TECHNICAL INFORMATION
7
4 87-900-869-01 (
D
)
Controller-to-Pump Connection
The pin configuration of the cable that connects the controller to the pump using a 0.5 mm
2
(AWG 20) wires is shown in
the following figure.
Controller-to-pump cable
RS 232 Communication Port
If the optional RS 232 communication port is installed,
refer to the following figures for pin connection and the
minimum connection configuration.
Communication serial port connections
The communication port mating connector is supplied
with the RS 232 PCB (AMP/Cannon or equivalent 15-
pin "D" type male connector).
The external cable (not supplied) between the host
computer and the controller doesn’t require any
crossed wires so that the signal are connected
correctly.
Minimum connection configuration
For example, the Transmit data signal from controller
(pin 2) must be connected to the host computer's
Receive data line (pin 3) and vice versa. Consult the
host computer's instruction manual for its serial port
connections.
NOTE
Varian cannot guarantee compliance with FCC
regulations for radiated emissions unless all external
wiring is shielded, with the shield being terminated to
the metal shroud on the 0-subconnector. The cable
should be secured to the connector with screws.