TECHNICAL INFORMATION
7
0 87-900-905-01 (
I
)
CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
100, 120, 220, 240 Vac, 1-phase
50 to 60 Hz
1400 VA maximum
Output for pump:
Voltage
Frequency
Power
water cooling
air cooling
54 Vac nominal
]
10%
633 Hz
]
2%
450 W maximum
450 W
300 W
Operating temperature
0 °C to +40 °C
Storage temperature
-20 °C to +70 °C
Fuse:
•
Mains
•
Heater Jacket (F4)
2 x T5A(slow blow) for 220 or 240 input Voltage
2 x T10A(slow blow) for 100 or 120 input Voltage
T3.15A (slow blow)
P1 optoisolator input
Minimum ON 3mA Maximum 5mA
J2 optoisolator output
24 Vdc, 60 mA
J7 output (fan, forepump coil, etc.)
120 Vac, 1A disregarding the mains
J11 purge flow connector
12 Vdc power supply output and signal input
J12 heater jacket connector
120 Vac 2A power supply output
In compliance with norms
EN 55011 class A group 1
EN 61010-1
IEC1000-4-2,1000-4-3, 1000-4-4
Auxiliary connectors
P1
J2
J7
J14, J13
External INPUT signals (pins)
OUTPUT signals (sockets)
Output fan Voltage and Valve, forepump socket
RS 232/422/485 connection (optional)
Interconnecting cables
Mains cable (3-wire, 3-meter long)
Pump cable (6-wire, 5-meter long)
Weight (both models)
13,5 kg (29.8 lbs)