TECHNICAL INFORMATION
39/50 87-901-004-01
(B)
POWER CONNECTION
Input power connector (24 Vdc type) 969-8995
The following figure shows the input power connector.
Mains 24V
Input power connector 969-8996
The following figure shows the input power connector.
CONTROLLER CONNECTION
P2 – Fan
The 24 Vdc, 100 mA maximum output voltage is present after
Switch ON of the controller.
This is a dedicated output for the optional
Turbo-V 81-AG pump cooling fan (see orderable parts table).
If a fan is already installed, use the available adapter cable to
connect to the pump.
P2 Fan Connector
P1 – Vent
P1 Vent Connector
This is a dedicated 24 Vdc connector for control of the optional
vent valve (see orderable parts table).
The vent valve can be driven by the controller automatically or
via a serial line.
If a 110 Vac-powered vent valve has already been installed,
this valve must be replaced with a new model (see orderable
parts table).