TECHNICAL INFORMATION
4
9
/
50
87-901-004-01 (
B
)
N.
Read/
Write
Data
Type
Description
Permitted
Values
110
R/W
L
Interlock type
(default = 1)
Impulse = 0
Continuous = 1
111
R/W
L
Analog out-
put type:
output volt-
age signal
proportional
to frequency
or power
0 = frequency
1 = power
(default = 0)
2 = Pump tem-
perature
3 = Pressure
117
R/W
N
Low Speed
frequency
(Hz)
1100 to
“Maximum ro-
tational
frequency”
(win 120)
(default =
1100)
120
R/W
N
Rotational
frequency
setting (Hz)
1100 to 1350
(default =
1350)
122
R/W
L
Set vent
valve on/off
0 = Closed
1 = Open
(default = 1)
125
R/W
L
Set the vent
valve
operation
Automatic = 0
(see note 1.)
On command =
1 (see note 2.)
126
R/W
N
Vent valve
opening
delay
(expressed in
0.2 sec)
0 to 65535
(corresponding
to 0 to 13107
sec)
(default = 15)
147
R/W
N
Vent open
time
See “vent
connector”
paragraph
0 = infinite
1 bit = 0.2 sec
1 to 65535
155
R
N
Power limit
applied
Read the
maximum
permitted
power
watt
157
R/W
N
Gas load
type
Select the
gas load to
the pump
Ø = N
2
1 = Ar
161
R/W
N
Pressure
Reading Con-
nection Factor
See “GAUGE
CONNEC-
TOR” for more
details
0 to 10
0 = 0
10 = 1 = N
2
162
R/W
A
R1 Set Point
Pressure
Valve
Valid if Win.
101 = 4
Format X.X
E. XX
Where X = 0
to 9
s = + or -
N.
Read/
Write
Data
Type
Description
Permitted
Values
163
R/W
N
Pressure unit
of measure
0 = mBar
1 = Pa
2 = Torr
167
R/W
A
Stop speed
reading
Activates /
deactivates
the pump
speed read-
ing after Stop
command
0 = disable
1 = enable
(default = 0)
200
R
N
Pump
current in
mA dc
201
R
N
Pump
voltage in
Vdc
202
R
N
Pump power
in W (pump
current x
pump volt-
age
203
R
N
Driving fre-
quency
in Hz
204
R
N
Pump
temperature
in °C
0 to 110
(60°C =
Threshold
Pump
Overtem-
perature)
205
R
N
Pump status
Stop = 0
Waiting intlk
= 1
Starting = 2
Auto-tuning =
3
Braking = 4
Normal = 5
Fail = 6
206
R
N
Error code
Bit descrip-
tion: see the
following fig-
ure
211
R
N
Controller
Heatsink
Temperature
(°C)
0 to 110
216
R
N
Controller Air
Temperature
(°C).
0 to 70
224
R
A
Pressure
reading
Format = X.X
E XX
226
R
N
Rotation
Frequency
(rpm)
300
R
N
Cycle time in
minutes (ze-
roed by the
reset com-
mand)
0 to 999999