TECHNICAL INFORMATION
64/74
87-901-010-01
EyeSys Mini-IMG gauge dimension
FRG-700 gauge dimension
J5 – IN-OUT
This connector carries all the input and output sig-
nals to remote control the TPS-compact.
It is a 15-pins D type connector; the available sig-
nals are detailed in the table, the following para-
graphs describe the signal characteristics and use.
PIN N.
SIGNAL NAME
INPUT/
OUTPUT
1 START/STOP
(+)
IN
2 START/STOP
(-)
IN
3 INTERLOCK
(+)
IN
4 INTERLOCK
(-)
IN
5
SPEED SETTING (+)
IN
6
SPEED SETTING (-)
IN
7
SOFT START (+)
IN
8
SOFT START (-)
IN
9 +24
Vdc
I
max
= 30 mA
OUT
10
NORMAL OUTPUT (relay)
V
max
= 125V
I
max
= 200 mA
OUT
11
PROGRAMMABLE SET POINT
OUT
12 NORMAL
OUTPUT
OUT
13 FAULT
OUTPUT
OUT
14 PROGRAMMABLE
ANALOG
SIGNAL (+)
OUT
15
•
GROUND
•
PROGRAMMABLE
ANALOG
SIGNAL (-)
OUT
Signal description
START/STOP
: input signal to start or stop the
pump. Without the supplied cover connector the
turbomolecular pump automatically starts, if you
want to stop the turbo-pump from outside, you
have to connect the START/STOP (+) signal to the
+24 Vdc pin and the START/STOP (-) signal to the
GROUND pin (this is the default configuration of
the cover connector).
INTERLOCK
: input signal to control the pump ro-
tation. With the supplied cover connector the IN-
TERLOCK (+) signal is connected to the +24 Vdc
pin and the INTERLOCK (-) signal to the
GROUND pin; in this condition the pump is
stopped.
SOFT START
: this input is used to provide a "soft
start" to the pump; in this condition the ramp-up
time could be up to 20 min.
SPEED SETTING
: PWM input signal to set the
pump speed. The PWM signal characteristics must
be the following:
•
frequency: 100 Hz +/-20%