Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-35
3.17 Using the Pendant for Parameter Setup
The service pendant is stored in the turret and is connected to Models 947/948/960 at
J1 in the turret for parameter changes, monitoring, or diagnostics. The pendant must be
connected to the 947/948/960 to operate any of the functions. Refer to Appendix G, “Leak
Detection Theory” for information on turret connectors.
The service pendant is used to set up and monitor:
❑
Reject set point
❑
Crossover pressures
❑
Auto sequence parameters
❑
Ion Source parameters
❑
System setup parameters (such as the value of the calibrated leak).
When the pendant is powered by the 947/948/960, the display indicates the model to
which the leak detector is connected. The keyboard is laid out in function keys and numeric
keys. See Figure 3-10.
Figure 3-10
Service Pendant Keypad
Several parameters relating to the operation and performance of the 947/948/960 Platform
are stored in EPROM. These parameters are viewed and changed via the service or rear
serial ports. Some parameters are set when the system is assembled and tested. Other
parameters should be changed when a gauge is changed or when the ion source or
preamplifier is changed. Some parameters are changed automatically during power up,
Auto Cal, or Auto Zero functions. See Appendix G, “Leak Detection Theory” for details.
Содержание 960
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 32: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual xxxii...
Страница 74: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 148: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 156: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 5 8...
Страница 194: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 202: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 216: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 229: ......