TECHNICAL INFORMATION 

be reduced later to optimize pumping and 
achieve the lowest vacuum pressure. 

5.  If started at 10

-4

 Torr (mbar) the voltage will 

start at approximately 800 volts and increase 
to full voltage as the pump starts operating. 
The current will start at several milliamps and 
slowly decrease to microamps or nanoamps 
as low vacuum pressure is achieved. 

6.  When first starting the pump, if the voltage de-

creases instead of increasing, reduce the vac-
uum pressure then start the ion pump again. 

7.  When the pump reaches its full operating volt-

age, you may close the roughing valve. 

8.  If the pump does not start after 30 minutes of 

pumping, see the section PUMP 
TROUBLESHOOTING. 

9.  Once the pump reaches its base pressure with 

stable voltage and current, the ion pump may 
be baked at high temperature if required per 
the section titled BAKEOUT OPERATION. 

10. To stop the ion pump, simply switch off the 

high voltage. The pump surfaces will continue 
to pump for a few minutes de

pending on the 

system pressure.

 

11. When venting the pump use clean, 

dr

y nitro-

gen. This will avoid water absorption on the 
pump surfaces and make subsequent pump 
do

wns

 easier

.

 

MAINTENANCE 

,

 

WARNING!

 

The high voltage present in the high voltage cable 
which connects the control unit to the ion pump, 
can cause severe injury or death. Before mounting 
the high voltage connector of the cable on the 
pump high voltage feedthrough, or before remov-
ing it, be sure the main power is removed from the 
control unit. 

 

Before removing the high voltage connector of the 
cable from the control unit, be sure the main power 
is removed from the control unit. Wait at least 10 
seconds after removing the main power from the 

control unit, to allow capacitors to discharge com-
pletely. 

VacIon 

Plus

 pumps are maintenance free. In case 

of life time expiry or premature failure of the pump, 
please contact your nearest Varian sales/service 
office for repair. 

The VacIon 

Plus

 20 pumps are designed with ex-

changeable high-voltage feedthrough and pump 
elements. 

Exchange of the high voltage feedthrough. 

   CAUTION 

The high voltage feedthrough contains a ceramic 
insulator that can be damaged if excessive force is 
applied in torque, in bending, or in tension. When 
installing or removing the feedthrough, the applied 
torque should not exceed 2 Nm. Rotate the 
feedthrough gently when making the threaded 
connection to avoid excessive force. 

1.  Remove the 6 bolts of the Mini-ConFlat flange 

connection (see Fig. 19). 

2.  Remove the cable connection disc. 

3.  Gently relieve the feedthrough from the metal 

gasket connection and turn the feedthrough 
counterclockwise until it is completely detached 
from the internal high voltage threaded connec-
tion. 

4.  Replace the feedthrough and the copper gas-

ket, making sure that the feedthrough is con-
nected to the pumps internal connector. Check 
that the feedthrough is not shorted. 

5. Bake out the pump while it is operating, and 

leave it to cool down; then verify that the base 
pressure is below 10

-9

 Torr (mbar). 

 

Fig. 19 - High voltage feedthrough connection 

 18 

87-900-106-01(B) 

Содержание 919-1115

Страница 1: ...del 919 1115 Model 919 1125 Model 919 1145 Model 919 1146 Model 919 1114 Model 919 1124 Model 919 1144 I MANUALE DI ISTRUZIONI D BEDIENUNGSHANDBUCH F NOTICE DE MODE D EMPLOI GB INSTRUCTION MANUAL 87 9...

Страница 2: ...VacIon Plus 20...

Страница 3: ......

Страница 4: ......

Страница 5: ...Vacuum evaluation 14 Short circuits 14 TYPICAL INSTALLATION 14 INLET FLANGE CONNECTION 15 CONTROL UNIT CONNECTION 16 SAFETY INTERLOCK 16 HEATER INSTALLATION 16 HEATER REPLACEMENT 17 BAKEOUT OPERATION...

Страница 6: ...engono informazioni importanti estrapolate dal te sto PREPARAZIONE PER L INSTALLAZIONE La pompa viene fornita in un imballo protettivo speciale se si presentano segni di danni che potrebbero essersi v...

Страница 7: ...zione all aria il deposito del composto di titanio assorbe vapore acqueo in con seguenza di ci al successivo avvio il riscaldamento della pom pa provoca il rilascio del vapore e di parte dell idrogeno...

Страница 8: ...hr besteht ACHTUNG Die Warnhinweise vor bestimmten Prozeduren machen den Bediener darauf aufmerksam da bei Nichteinhaltung Sch den am Ger t entstehen k nnen W hrend des Betriebes soll die Umgebungstem...

Страница 9: ...stromwert der Steuereinheit ann hert weist auf eine Leckstelle an der Pumpe und am System hin Ein zeitweiliger Anstieg des Vorvakuumdruckes ist w hrend der Anlaufphase nor mal 4 W hrend des Anlaufes d...

Страница 10: ...fonctionne ment pour respecter les sp cifications techniques d clar es la temp rature ambiante doit tre comprise entre 0 C et 85 C ATTENTION Les messages d attention apparaissent avant certaines proc...

Страница 11: ...400 V Une valeur de courant proche du courant de court circuit de l unit de contr le r v le l existence d une fuite dans la pompe et dans le syst me Au cours de la phase de d marrage un accroissement...

Страница 12: ...in explosive environments or those with a high fire risk Dur ing operation to obtain the specified operation the ambient temperature must be between 0 C and 85 C CAUTION The caution messages are displ...

Страница 13: ...umped hydrogen thus the starting time may be lengthened CAUTION Do not put any electronic device near the pump otherwise the magnetic field around it may cause a device malfunctioning MAINTENANCE Oper...

Страница 14: ...ty high voltage supply is required to operate Diode and Noble Diode pumps A nega tive polarity high voltage supply is required to op erate StarCell pump since the anode is grounded and the cathode is...

Страница 15: ...ate the VacIon Plus 20 pump it normally requires an amount of gas equal to 0 6 Torr litres mbar litres Consequently pumps can operate for extended periods of time at low pressures in the non saturated...

Страница 16: ...TECHNICAL INFORMATION VACION PLUS 20 NOBLE DIODE Fig 6 Pumping speed vs pressure for Nitrogen Fig 7 Pumping speed vs pressure for Argon Fig 8 Pressure vs current diagram 11 87 900 106 01 B...

Страница 17: ...0 DIODE Fig 9 Pumping speed vs pressure for Nitrogen Fig 10 Pressure vs current diagram OUTLINE DRAWING The following figure shows the outline drawing for the VacIon Plus pump Fig 11 VacIon Plus pump...

Страница 18: ...ne of the pump and in the plane of the flange as a function of distance from the pump are shown in Figs 13 14 and 15 Fig 12 VacIon Plus pump axis identification Fig 13 Stray magnetic field along X axi...

Страница 19: ...or an indication of one of the following conditions If the pump is free of leaks and is at a low pressure the pressure indication shall within 10 minutes fall to or below the 10 8 Torr mbar range as t...

Страница 20: ...nal vacuum using pliers to open the copper tube pinch off CAUTION Do not open the pinch off seal with a saw or grinder These methods will cause metal particles to be drawn into the pump by the inrushi...

Страница 21: ...or blanked off WARNING Before removing the high voltage connector of the cable from the control unit be sure the main power is removed from the control unit Wait at least 10 seconds after removing the...

Страница 22: ...the pump and system cool down to room temperature a drop in pressure should be ob served Note that the ion pump can be also baked when switched off into an external turbo pump through a bakeable isola...

Страница 23: ...it be sure the main power is removed from the control unit Before removing the high voltage connector of the cable from the control unit be sure the main power is removed from the control unit Wait at...

Страница 24: ...emis sion current high potting should be performed High potting is the term used to describe the ap plication of higher than normal operating voltage 10 12 KV 20 50 mA for the purpose of burning off...

Страница 25: ...and pump Subse quent reduction in pressure de pends on the rate of depletion of this vapour Heavy hydrocarbons are most troublesome because of their relative low vapour pressure and they are very diff...

Страница 26: ...Ferrite magnets 919 0040 919 0039 919 0035 Samarium Cobalt magnets 919 0036 Heaters 120V 220V 919 0044 919 0043 Copper gasket for 2 3 4 inlet flange 10 pack individually sealed FG 0275 CI Nut and bolt...

Страница 27: ...0 Vac 2 Neg 120 Vac 1 Pos 120 Vac 2 Pos 230 Vac 1 Neg 230 Vac 2 Neg 230 Vac 1 Pos 230 Vac 2 Pos 929 7008 929 7009 929 7010 929 7011 929 7012 929 7013 929 7014 929 7015 MidiVac 120 Vac with single outp...

Страница 28: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 29: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 30: ...1263 Toll Free 0120 655 040 Korea Varian Technologies Korea Ltd Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 2 3452 2452 Fax 82 2 3452 2451 Toll Free 080 222 2452 Mexico...

Отзывы: