WARNING:
SHOCK HAZARD
Dangerous voltages exposed when cover is removed. Unplug power cord.
3.
Be sure the transfer line is cool.
Retract the transfer line by grabbing the
front nose and turning counterclockwise while pulling out. A mild amount of
force may be needed to release residual vacuum. If the transfer line does not
pull out, reopen the vent to be sure the analyzer is at atmospheric pressure.
If the transfer line is still difficult to retract, try using the transfer line tool or a
3/16” wrench to twist the end of the transfer line counterclockwise as you
retract it. After the transfer line has been fully retracted, lock it into position by
turning it clockwise.