Switching from Liquid to Gaseous CI Reagent
Operation
To switch from the Liquid CI Inlet back to a pressurized CI Gas (such as
methane), the CI Gas line may be Reinstalled without removing the liquid CI inlet
assembly.
1. Loosen the 2 screws that attach the liquid CI inlet restrictor to the back of the
instrument. Also, loosen the 2 screws that attach the L-bracket to the liquid
CI inlet block.
2. Remove the liquid CI restrictor end that inserts into the back of the
instrument; rotate the restrictor out of the way.
3. Install the 4 mL/min CI Gas restrictor (03-930597-01) between the gas supply
and the CI shutoff block, below the L-bracket.
4. Tighten all screws.
5. Adjust CI reagent as described in the User Manual.
If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.