Acquiring OFN
1. Use the Inject Single Sample option.
2. Use a lower air gap to avoid losing sample through the tip of the needle.
Ensure the syringe plunger is not forced out of the syringe barrel if using a
Pressure Pulse.
3. Strike the injector auto switch with the syringe barrel. This ensures that the
split valve has changed from open to close.
4. Make a 1 µL manual injection of OFN
Reviewing the Data
1. View a data file MS Data Review.
2. Enter 272 in the
Ions:
field of the MSDR window to display a single ion
chromatogram of the m/z 272 ion signal.