Adjusting Cal Gas
To adjust the cal gas:
1. Open the MS door and ensure that the Cal Gas vial is full.
2. Click
Adjust Cal Gas
.
3. Adjust the Cal Gas needle valve behind the MS front door until the display
shows that the Cal Gas pressure is OK.
4. Click
Save Results
when finished.
NOTE: The Cal Gas adjustment varies with differences in auto tune settings,
filament emission current, system sensitivity, and system background levels.
Check this adjustment after auto tuning, editing the filament emission current, or
when system sensitivity, or background levels change.
03-954143-55:Rev. 1
15 of 33