4. Remove the turbo pump shipping restraint from the MS.
5. Remove the shipping restraints from the GC including the metal dowel at the
rear of the GC.
6. Instruct the customer to keep the removed restraints, in case the system
needs to be transported.
7. Connect the foreline hose to the foreline pump. Connect the power cord of
the foreline pump to the LINE VOLTAGE PUMP ONLY connector at the rear
of the MS. To minimize vibrations, put the foreline pump on the floor not on
the bench.
8. Connect the power cords of the MS, the GC, and the PC into the appropriate
outlets.
9. Connect the GC to the MS as follows:
•
431-GC: Connect the sync cable to the back of the GC at J2 sync
analog out and to the back of the MS at J43 Remote Option.
•
450-GC: Connect the sync cable under the GC top cover at J401
sync analog out and to the back of the MS at J43 Remote Option.
J2 Sync/Analog Out
J2 Sync/Analog Out on back of 431-GC