background image

Table 8 Maximum Power Draws for Varian Modules 

Component 

100-115 Vac 

(Amperes) 

200-240 Vac 

(Amperes) 

Mass Spectrometer 

12 

LC/MS Foreline pumps (2 ea. @ 220V) 

N/A 

20 

ProStar 210/215 SDM 

2.25 

1.125 

ProStar 220, 230, 240 SDM 

2.25 

1.125 

PrepStar SD-1 SDM 

3.0 

1.5 

PrepStar SD-2 (220V only) SDM 

3.3 

1.65 

PrepStar 218 SDM 

4.6 

2.3 

ProStar 310 UV-Vis 

2.0 

ProStar 320 UV-Vis 

1.3 

0.75 

ProStar 325 Photodiode Array 

1.1 

0.55 

ProStar 330 Photodiode Array 

1.7 

0.9 

ProStar 335 Photodiode Array 

1.1 

0.55 

ProStar 340 UV-Vis 

0.9 

0.45 

ProStar 345 UV-Vis 

1.7 

0.9 

ProStar 363 Fluorescence 

3.0 

1.5 

ProStar 400 AutoSampler 

1.7 

0.85 

ProStar 410 AutoSampler 

1.25 

0.625 

ProStar 420 AutoSampler 

1.7 

0.9 

ProStar 430 AutoSampler 

2.1 

1.05 

ProStar 701 Fraction Collector 

0.6 

0.3 

ProStar 704 Fraction Collector 

0.2 

0.1 

CP-3800 Gas Chromatograph 

20 

10 

CP-8400 AutoSampler  

0.5 

0.25 

Combi PAL 

0.7 

1.5 

Computer  

1.5 

Monitor  

1.5 

Laser Printer 

3-4 

1.5-2 

Never plug the mass spectrometer and the chromatograph into the same power source; 

otherwise, you may overload the power source. Never use the free outlet on any of the power 

sources for equipment drawing more than 2A. 
The quality of the power supplied to your LC/MS or GC/MS is very important. The power must 

be 100-130 Vac, 60 Hz 

±

3 Hz (200-260 Vac, 50 Hz 

±

3 Hz outside North America), and 200-240 

Vac, 60 Hz 

±

3 Hz for LC/MS (50 Hz 

±

3 Hz for outside North America), and it must be stable, 

i.e., it must be free of fluctuations due to slow changes in the average voltage or to changes 

resulting from surges, sags, or transients. 

  Slow average changes are gradual, long-term changes in the average root mean 

square (RMS) voltage level, with typical durations greater than 2 seconds. 

  Sags and surges are sudden changes in average RMS voltage level, with typical 

durations between 50 µsec and 2 seconds. 

  Transients (or impulses) are brief voltage excursions of up to several thousand volts 

with durations of less than 50 µsec. 

03-914961-00:6 

23

 

of 24 

1200/1200L MS Pre-installation Instructions

 

Содержание 1200 GC/MS

Страница 1: ...e relative humidity can be maintained between 40 80 The lab is free of particulate matter The system workbench is free from vibrations The system area is free from electromagnetic disruption and elect...

Страница 2: ...in this manual could cause failures of many types The repair of such failures is specifically excluded from the Warranty and Service contract conditions CAUTION All phases of the installation site pre...

Страница 3: ...densation Operating the mass spectrometer at very low humidity may result in the accumulation and discharge of static electricity shortening the life of electronic components Operating the system at h...

Страница 4: ...hrough doors CAUTION The 1200 MS foreline pump and gas chromatograph are heavy items that require at least two fit people for lifting and moving safely Space and Load Requirements The recommended layo...

Страница 5: ...on devices or test equipment we recommend a separate dedicated power source for their operation Never plug the mass spectrometer and the chromatograph into the same power source or you may overload th...

Страница 6: ...ferably flame treated or solvent washed Ammonia 5 0 Research Grade 99 999 purity A two stage 0 15 psi 0 1 bar pressure regulator that has a stainless steel diaphragm is recommended The output pressure...

Страница 7: ...ation Your GC Accessory Kit includes a GC MS Gas Clean Oxygen Moisture filter CP 17973 and filter base This easy to remove gas filter cartridge combines three highly adsorptive materials in one filter...

Страница 8: ...hin 10 ft 3m of the rear of the 1200L NOTE See Appendix 1 for exact module specifications Power Requirements 1200L GC MS The 1200L GC MS system requires one dedicated duplex single phase power source...

Страница 9: ...1200L GC MS CI Reagent Gases The CI mode for the 1200L MS typically uses methane isobutane or ammonia The amount of gas consumed during CI operation is approximately 1 to 2 mL min The CI reagent gas...

Страница 10: ...ate precautions when using Anhydrous Ammonia and for compliance with all governmental regulations It is the operator s responsibility to understand and adhere to all safe laboratory practices concerni...

Страница 11: ...sed with your filter 1200L LC MS Entrance Before arranging for delivery of the 1200L LC MS to your facility please make sure that all passages to the site of installation are at least 36 in 91 5 cm wi...

Страница 12: ...rces with earth grounds hard wired to the main power panel ground and one 230V 20A power source One of these fourplex power sources is for the mass spectrometer computer and monitor the other fourplex...

Страница 13: ...he CEE 7 7 plug and outlet are shown in Appendix 3 Figure 5 c Gas Requirements 1200L LC MS Nitrogen A Nitrogen gas supply capable of providing up to 12L min of gas regulated at 80 psi is required as a...

Страница 14: ...ld always be kept to a minimum Semiprep and Preparative systems have other requirements limitations with regard to interconnect tubing Typically a system using flow rates up to 200 mL min will require...

Страница 15: ...power sources with earth grounds hard wired to the main power panel ground for the HPLC MS and computer and one 230V 20A power source for the foreline pumps If you have additional sample preparation d...

Страница 16: ...g and corresponding outlet are shown in Appendix 3 Figure 5 a NEMA 5 20P plugs are rated at 20A and 120 Vac The power cables for the computer monitor and printer are approximately 78 in 2m long They a...

Страница 17: ...the line be as short as possible Methane 5 0 Research Grade 99 999 purity A two stage 0 15 psi 0 1 bar pressure regulator that has a stainless steel diaphragm is recommended The output pressure to the...

Страница 18: ...Carrier Gas Helium Helium is required as a carrier gas for the GC with a minimum 99 999 ultra high purity and less than 1 0 ppm each of water oxygen and total hydrocarbons The minimum requirement is o...

Страница 19: ...ing or installation of modules to existing HPLC systems For most analytical HPLC systems flow rates up to 10 mL min 1 16 tubing with an internal diameter ID of 0 005 can be used throughout However 0 0...

Страница 20: ...with the observations Contact the appropriate Varian office to report the damage If no external damage is apparent receiving documents should be signed Received but not inspected to indicate that the...

Страница 21: ...ted and you have accepted the conditions of delivery Spare Parts The 1200 1200L Hardware Operation Manual provides a list of spare parts for routine operation Preventive Maintenance The customer is re...

Страница 22: ...325 Photodiode Array Detector 8 3 21 2 11 7 29 6 18 7 47 5 34 15 5 ProStar 330 Photodiode Array Detector 9 3 23 5 10 2 26 20 50 35 16 ProStar 335 Photodiode Array Detector 8 3 21 2 11 7 29 6 18 7 47...

Страница 23: ...to 12 in 15 to 30 cm Allow vertical clearance above the instrument Dissipate heat and allow for routine maintenance 30 in 76 cm Allow additional bench space beside the Workstation keyboard Enable mous...

Страница 24: ...5 Laser Printer 3 4 1 5 2 Never plug the mass spectrometer and the chromatograph into the same power source otherwise you may overload the power source Never use the free outlet on any of the power s...

Страница 25: ...orten device lifetime It is important to establish the quality of the line power in your laboratory prior to installation of the 1200L LC MS and or GC MS Occasionally line power sources may be of unac...

Отзывы: