User Manual: UM50566-EN
Optica 30/40/50 series
Revision: 3.0
Date of release: 2019-01-09
31/36
www.vareximaging.com
SID precision of measurement with
retractable tape
(EN60601-1-3: §8.5.2)
< 2 % SID
Leakage radiation
(At 1 m, X-ray beam = 150 kVp, 4 mA;
EN60601-1-3: §12.4)
≤ 0.5 mGy/h
Collimator rotation (+/- 90°)
Optional with rotating flange
Remote control interface
Yes (via CAN bus)
Maximum load for accessory rails
70 N
Weight
9.0 kg
Size
230 x 320 x 210 mm (w x d x h)
Operation environment
Ambient temperature: 10 – 40 °C
Relative humidity:
30 – 95 %, non-condensing
Atm. Pressure:
700 – 1060 hPa
Storage
Ambient temperature: -20 – +70 °C
Relative humidity:
30 – 95 %, non-condensing
Atm. Pressure:
700 – 1060 hPa
6.1.
Product label
The figure below shows the information provided on the product label. Where “Model: Optica XX”
represents the model name of the collimator.
Product label for Optica 40
Medical Device
Anholtseweg 44
7091 HB Dinxperlo
The Netherlands
www.vareximaging.com
24 VDC / 80VA
0120
710383
OPTICA 40
XXXXX-NN-LLL
XXXXXXXXXXXX
YYYY-MM-DD
Model:
REF
SN
Min. inherent filtration
Al. Equiv.: 1.2mmAl / 75 kV
X-ray rating up to: 150 kVp IP20
(01)0XXXXXXXXXXXX(11)XXXXXX(21)WOXXXXXX/E-XX
This product complies with the requirements of 21 CFR Subchapter J for radiation emitting products in effect on the date of manufacture.
This product complies with 21 CFR 1040.10 and 1040. 11 except for deviations pursuant to Laser Notice No. 50, dated July 26, 2001
Only