Figure 39 Gain Calibration Complete
Note
Gain calibration should be performed at regular intervals:
•
At least, every twelve (12) months.
•
If the central beam of the X-ray source has been moved relative to the
X-ray detector.
•
If the X-ray tube is replaced.
•
If the X-ray detector is dropped.