background image

2.4 Intensité (courant alternatif) 

plage 

résolution 

précision 

600µA 

0.1µA 

+/- (1.2% + 3) 

6000µA 

1µA 

+/- (1.2% + 3) 

60mA 

0.01mA 

+/- (1.2% + 3) 

600mA 

0.1mA 

+/- (1.2% + 3) 

6A 

0.001A 

+/- (1.5% + 5) 

10A 

0.01A 

+/- (1.5% + 5) 

Protection : 
Plage mA : fusible F1 

ɸ6x32mm F 600mA H 600V 

Plage 10A : fusible F2 

ɸ6x25mm, F 10 A H 600V 

 

2.5 Résistance 

Plage 

résolution 

précision 

6

00Ω 

0.1Ω 

+/- (1.0% + 2) 

6

kΩ 

0.001kΩ 

+/- (0.8% + 2) 

6

0kΩ 

0.01kΩ 

+/- (0.8% + 2) 

6

00kΩ 

0.1kΩ 

+/- (0.8% + 2) 

6

0MΩ 

0.01MΩ 

+/- (2.0% + 5) 

Protection : 
600Vrms 
 

2.6 Capacitance 

Plage 

résolution 

précision 

9.999nF 

0.001nF 

+/- (4.0% + 10) 

99.99nF 

0.01nF 

+/- (4.0% + 5) 

999.9nF 

0.1nF 

+/- (4.0% + 5) 

9.999µF 

0.001µF 

+/- (4.0% + 5) 

99.99µF 

0.01µF 

+/- (4.0% + 5) 

999.9µF 

0.1µF 

+/- (4.0% + 5) 

9.999mF 

0.001mF 

+/- 10% 

Protection : 
600V-PTC 
Test 

capacitance ≤ 200nF, adapté au mode REL 

 

2.7 Température 

Plage 

résolution 

précision 

°C 

-40 ~ 
1000°C 

-40~40°C 

1°C 

+/- 4°C 

>40~500°C 

+/- (1.0% + 4) 

>500~1000°C 

+/- (2.0% + 4) 

°F 

-40 ~ 
1832°F 

-40~104°F 

1°F 

+/- 5°F 

>104~932°F 

+/- (1.5% + 5) 

>932~1832°F 

+/- (2.5% + 5) 

Protection : 
600V 
K-type thermocouple type K est seulement possible pour les 
températures inférieures à 250°C/482°F 
 
 
 

2.4 AC current 

range 

resolution 

accuracy 

600µA 

0.1µA 

+/- (1.2% + 3) 

6000µA 

1µA 

+/- (1.2% + 3) 

60mA 

0.01mA 

+/- (1.2% + 3) 

600mA 

0.1mA 

+/- (1.2% + 3) 

6A 

0.001A 

+/- (1.5% + 5) 

10A 

0.01A 

+/- (1.5% + 5) 

Protection : 
mA range : fuse F1 

ɸ6x32mm F 600mA H 600V 

10A range : fuse F2 

ɸ6x25mm, F 10 A H 600V 

 

2.5 Resistance 

range 

resolution 

accuracy 

6

00Ω 

0.1Ω 

+/- (1.0% + 2) 

6

kΩ 

0.001kΩ 

+/- (0.8% + 2) 

6

0kΩ 

0.01kΩ 

+/- (0.8% + 2) 

6

00kΩ 

0.1kΩ 

+/- (0.8% + 2) 

6

0MΩ 

0.01MΩ 

+/- (2.0% + 5) 

Protection : 
600Vrms 
 

2.6 Capacitance 

range 

resolution 

accuracy 

9.999nF 

0.001nF 

+/- (4.0% + 10) 

99.99nF 

0.01nF 

+/- (4.0% + 5) 

999.9nF 

0.1nF 

+/- (4.0% + 5) 

9.999µF 

0.001µF 

+/- (4.0% + 5) 

99.99µF 

0.01µF 

+/- (4.0% + 5) 

999.9µF 

0.1µF 

+/- (4.0% + 5) 

9.999mF 

0.001mF 

+/- 10% 

Protection : 
600V-PTC 
Test capacitance ≤ 200nF, adapt REL mode 
 

2.7 Temperature 

range 

resolution 

accuracy 

°C 

-40 ~ 
1000°C 

-40~40°C 

1°C 

+/- 4°C 

>40~500°C 

+/- (1.0% + 4) 

>500~1000°C 

+/- (2.0% + 4) 

°F 

-40 ~ 
1832°F 

-40~104°F 

1°F 

+/- 5°F 

>104~932°F 

+/- (1.5% + 5) 

>932~1832°F 

+/- (2.5% + 5) 

Protection : 
600V 
K-type thermocouple is only applicable for temperature less than 
250°C/482°F 
 

3. MAINTENANCE 

 
Avant de remplacer la batterie, débranchez les pointes de test 
et éteignez le multimètre. Dévisser les vis du capot de la 
batterie. Spécification de la batterie : AAA 1.5V x2. 
 
Pour remplacer les fusibles, dévissez les vis à l’arrière de 
l’appareil. Spécification des fusibles : 
F1 

ɸ6x32mm, F 600mA H 600V 

F2 

ɸ6x25mm, F 10 A H 600V 

3. MAINTENANCE 

 
Before replacing the battery, remove the test leads and switch off 
the multimeter. Take off the screw on the battery cover. Battery 
specifications : AAA 1.5V x2. 
 
To replace fuse, take off the screws on the back of the 
multimeter. Fuse specifications : 
F1 

ɸ6x32mm, F 600mA H 600V 

F2 ɸ6x25mm, F 10 A H 600V 

 

Отзывы: