Page 30
3001002 Rev. T
Appendix
Recommended separation distances between portable and mobile RF communications equipment and the Precision
Flow®
Max Output Power
(Watts)
Separation (m) 150kHz
to 80MHz
D=(3.5/V1)(√P)
Separation (m) 80 to
800MHz
D=(3.5/E1)(√P)
Separation (m) 800MHz to
2.5GHz
D=(7/E1)(√P)
0.01
.1166
.1166
.2333
0.1
.3689
.3689
.7378
1
1.1666
1.1666
2.3333
10
3.6893
3.6893
7.3786
100
11.6666
11.6666
23.3333
Guidance and manufacturer’s declaration - electromagnetic emissions
The Precision Flow® is intended for use in the electromagnetic environment specified below.
The customer or the user of the Precision Flow® should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment- guidance
RF emissions
CISPR 11
Group 1
The Precision Flow® uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby equipment.
RF emissions
CISPR 11
Class B
The Precision Flow® is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes, provided the following
warning is heeded:
Warning:
This equipment/system is intended for use by healthcare pro-
fessionals only. This equipment/system may cause radio interference or
may disrupt the operation of nearby equipment. It may be necessary to
take mitigating measures, such as re-orienting or relocating the Precision
Flow® or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration- electromagnetic immunity
IEC 60601-1-2:2001+A1:2004
EN60601-1-2:2001
AS/NZ3200.1.2:2005
Sub Test
Passed Parameters
Electro-Static Discharge
EN 61000-4-2:1995,+A1:1998,
+A2:2001
±6kV Contact discharge
±8kV Air discharge
Radiated RF Susceptibility
EN 61000-4-3:2002
80- 2500MHz @ 3 V/m,
1kHz AM 80% modulation
Electrical Fast Transients
EN 61000-4-4:2004
±5kV AC mains,
Surges
EN 61000-4-5:1995, +A1:2001
±0.5,1kV Line to line
±0.5, 1 2kV Line to protected earth
Line Conducted RF Susceptibility
EN 61000-4-6: 1996, +A1:2001
0.15-80MHz @ 3Vrms,
1kHz AM 80% modulation
Power Frequency Magnetics
EN 61000-4-8: 1993, +A1:2001
3A/m @ 50/60Hz PASSED
Voltage Dips and Dropouts
EN 61000-4-11: 2004
Per Standard