Page 14
3001002 Rev. T
Section 7
Setting up
7-5-2. Connect the delivery tube to the disposable water path as shown.
Press firmly into place.
NOTE
: If the door does not close easily, check that the cartridge is installed correctly
and the disposable water path is fully inserted into the docking station.
CAUTION
: Do not remove the disposable patient circuit while the unit is operating
7-6. Inserting disposable patient circuit:
7-6-1. Open door to expose the docking station.
7-6-2. Hold disposable patient circuit by its handle,
with the delivery tube downward as shown.
7-6-3. Slide disposable patient circuit downward into
the docking station until it stops.
7-6-4. Press down firmly so there is no gap between the
bottom of the Disposable Water Path and the docking
station floor.
7-6-5. Close door.