9. H YGIEN E
purpose, never metal brushes or
steel wool. If necessary, repeat this
procedure several times until no
remaining residue is visible on the
parts.
Remove the parts from the soapy
water and rinse them for at least 1
min. under running water.
The Filling Chamber parts, Tube
Flexer and Mouthpiece can also be
cleaned in the dishwasher using a
household dishwasher detergent.
Place the individual parts in a
suitable place in the dishwasher.
Small parts (e.g. Filling Chamber
Rings or Mouthpieces) can get lost,
or could damage the dishwasher,
if the parts are not appropriately
placed in the dishwasher. Surface
changes can occur in the Filling
Chamber Cylinder and Dosing
Capsule Adapter in a dishwasher.
After cleaning, allow the liquid
inside to run out completely and
shake any remaining water out of
the parts. Avoid any contact with
unclean surfaces.
Note: If the tap water is very hard, re-
sulting in the formation of a white film
on the surface of the parts, we recom-
mend replacing the Filling Chamber
and Tube Kit more frequently.
Possible alternative:
alternatively, all
the Filling Chamber parts, as well as
the Tube Flexer, and Mouthpiece, can
be cleaned in isopropyl alcohol.
Manual cleaningwith isopropyl alcohol (70 %)
or detergent
Warning!
Isopropyl alcohol is flam
-
mable and may easily ignite.
The safety instructions fur-
nished by the manufactu-
rer of the isopropyl alcohol
must be observed!
Do not immerse any plastic
parts in isopropyl alcohol for
more than half an hour. Pro-
longed contact with isopropyl alcohol
can lead to color changes or embrittle-
ment of the plastic parts. The Herb Mill
and Magazin must not be placed in
isopropyl alcohol.