Guidance and Manufacturer‘s Declaration – Electromagnetic Emis-
sions
The
MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
Class A
Complies
The
MIGHTY MEDIC Vaporizer uses
RF energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The
MIGHTY MEDIC Vaporizer is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Electromagnetic environment - guidance
Emissions test
Compliance
10. THE
MIGHTY MEDIC VAPORIZER
10.5. Information on Electromagnetic Compatibility
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic envi-
ronment – guidance
Guidance and Manufacturer’s Declaration – Electromagnetic Immu-
nity
The
MIGHTY MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MIGHTY MED-
IC Vaporizer should assure that it is used in such an environment.
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for power
supply lines
not applicable:
input/output
lines are
non-existent
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.
Mains power quality
should be that of a
typical residential,
commercial or hospital
environment.
10. THE
MIGHTY MEDIC VAPORIZER