background image

35 

34 

Turn the Cooling Unit with a 90° ro-
tation counterclockwise in order to 
remove it from the 

MIGHTY MEDIC 

Vaporizer. 
Pull out the Mouthpiece from the Cool-
ing Unit Cap, rotate slightly back and 
forth while doing this, and pull off the 
Mouthpiece Seal Ring. 

9. DISASSEMBLY AND ASSEMBLY OF THE   

MIGHTY MEDIC COOLING UNIT

Pulling the Mouthpiece from the Cooling Unit 
Cap 

Opening the Cap Lock 

Open the Cap Lock by pulling it rear-
wards. 

Pull the Cooling Unit Cap from the Cooling 
Unit Base 

If the Cooling Unit must be im-
mersed into alcohol (ethanol) 
the Cap Lock only has to be 

removed from the Cooling Unit. Ex-
tended contact with alcohol can lead 
to whitening of the Cap Lock Label or 
disintegration of the glue. 

Removal of the Cap Lock 

Separate the Cooling Unit Cap from 
the Cooling Unit Base by pulling off 
the cap upwards.

Remove the Cap Lock from the cap by 
pulling up at one side.

Pressing out the upper Screen

Push the upper Screen downwards 
out of the Filling Chamber Cap from 
above by using the Filling Chamber 
Tool supplied. 

Reassemble in reverse order.

When assembling, make sure 
that Seal Rings are properly 
seated. The Mouthpiece Seal 

Ring, especially, must sit completely 
in the cap and be no longer visible. 
The upper Screen must be inserted 
carefully into the groove provided. 
Disregarding this may allow plant 
particles to enter the Cooling Unit, so 
blocking this or be inhaled. 
When using the Dosing Capsules dur-
ing operation, the Lower Screen will 
normally not get polluted. If the screen 
does have to be exchanged one day, 
proceed as follows: 
Remove the lower Filling Chamber 
Screen out of the Filling Chamber by 
using a pair of tweezers or a pair of 
pointed pliers. Insert a new Screen 
while taking care that the edges of the 
Screen seat into the groove at the bot-
tom of the Filling Chamber. 

9. DISASSEMBLY AND ASSEMBLY OF THE   

MIGHTY MEDIC COOLING UNIT

Removal of the large Base Seal Ring

Remove the large and small Base 
Seal Rings. Lever out the large Seal 
Ring from the recess in the notch us-
ing the Filling Chamber Tool. 

Содержание MIGHTY MEDIC

Страница 1: ...INSTRUCTIONS FOR USE Read carefully before using the device and save these instuctions...

Страница 2: ...GHTY MEDIC Vaporizer 17 5 1 Unpacking 17 5 2 Power Connection 17 5 3 Application and Operation 18 5 4 Charging the Battery 18 5 5 Heating Up 20 5 6 Temperature Setting 21 5 7 Reset to Default Settings...

Страница 3: ...zer 36 10 1 Cleaning of the MIGHTY MEDIC Vaporizer and the Power Adapter 36 10 2 Care and Maintenance Storage Disposal 36 10 3 Troubleshooting Periodic Safety Checks 37 10 4 Technical Specifications 3...

Страница 4: ...have an average expected operating life of approx 500 operating hours approx 2 years For exchanging the batteries please contact our Service Center Service Contact in case of technical problems or qu...

Страница 5: ...2005 The product may not be disposed of in normal household waste The X on the trash can indicates the necessity of disposing of this product separately The device contains a built in rechargeable lit...

Страница 6: ...Do not wind the power adapter cord tightly nor pull it over sharp edges and never crush or kink it If the power adapter cord is damaged please contact our Service Center to have it replaced Never try...

Страница 7: ...s The Filling Chamber is made of metal do not touch it when it is hot The Dosing Capsule is made of metal do not touch it when it is hot Before cleaning disconnect the power adapter from the socket Th...

Страница 8: ...ychotropic mind altering effect Thus there is a potential for abuse which governments in nearly all coun tries of the world counteract with cor responding laws Use only medicinal hemp flowers cannabis...

Страница 9: ...e Keep the Instructions for Use for future ref erence Remove the MIGHTY MEDIC Vapor izer and the accessories from the box Immediately after unpacking check whether the Vaporizer the accesso ries the p...

Страница 10: ...e frame and all six fields are visible The number of status fields visible depends on the extent of the battery charge Check the housings of the MIGHTY MEDIC Vaporizer and the power adapter for signs...

Страница 11: ...on If the basic temperature can no longer provide adequate evaporation the booster temperature can be activated by means of a double click on the or ange colored ON OFF switch Once the booster tempera...

Страница 12: ...ules by the nursing staff Inserting the Dosing Capsules into the MIGHTY MEDIC Vaporizer is con siderably easier than filling the Filling Chamber with ground hemp flowers 6 1 Filling the Dosing Capsule...

Страница 13: ...or any other cannabis that has not been sub jected to recognised quality controls The amount of active ingredients in these products is generally unknown may fluctuate considerably and they often cont...

Страница 14: ...the MIGHTY MEDIC Vapor izer has reached the set temperature 180 C 356 F set and actual tem peratures are the same also signaled by a short double vibration and the Mouthpiece with Lip Piece is folded...

Страница 15: ...be used 8 HYGIENE Material Resistance of the MIGHTY MEDIC Vaporizer Please ensure that the Lip Piece the Cooling Unit with Mouthpiece and the MIGHTY MEDIC Vaporizer do not come in contact with the fol...

Страница 16: ...or brittle Lip Pieces are not to be used 8 2 Cooling Unit with Mouthpiece When the vapors are cooled it is in evitable that a fraction is condensed on the internal surfaces of the Cool ing Unit The co...

Страница 17: ...iece more frequently As an alternative to warm water with wash up liquid all parts of the Cool ing unit with Mouthpiece exception Cap Lock can be cleaned with etha nol Manual cleaning of the disassemb...

Страница 18: ...h the upper Screen downwards out of the Filling Chamber Cap from above by using the Filling Chamber Tool supplied Reassemble in reverse order When assembling make sure that Seal Rings are properly sea...

Страница 19: ...alified persons In Case of Problems If the power adapter has been in use unplug it at the socket and ensure that no one can connect the MIGHTY MEDIC Vaporizer to the mains supply again without this be...

Страница 20: ...o be documented If the device is not functioning prop erly or fails any of the above tests the device has to be repaired 10 4 Technical Specifications Power adapter Manufacturer Mean Well types GSM36E...

Страница 21: ...on on Electromagnetic Compatibility Immunity test IEC 60601 test level Compliance level Electromagnetic envi ronment guidance Guidance and Manufacturer s Declaration Electromagnetic Immu nity The MIGH...

Страница 22: ...test level 10 THE MIGHTY MEDIC VAPORIZER Immunity test IEC 60601 test level Compliance level Electromagnetic envi ronment guidance Guidance and Manufacturer s Declaration Electromagnetic Immu nity The...

Страница 23: ...con sidered If the measured field strength in the location in which the MIGHTY MEDIC Vaporizer is used exceeds the applicable RF compliance level above the MIGHTY MEDIC Vaporizer should be observed t...

Страница 24: ...regular mesh width REF 03 11 MM MIGHTY MEDIC Normal Screen Set 6 pcs Lower Screen with regular mesh width REF 03 12 MM MIGHTY MEDIC Coarse Screen Set 6 pcs Upper Screen with coarse mesh width REF 04 0...

Страница 25: ...users Normal wear or tear to the products is not con sidered a defect under warranty law If the customer has commissioned third parties with completing maintenance or repair of the product or if the c...

Страница 26: ...imated costs and obtained cus tomer approval Copyright This document is copyright protected and may not be used either in whole or in part pursuant to Sections 15 et seqq UrhG German Copyright Act wit...

Страница 27: ...AL 30 210 ENG 08 2017 Subject to alterations All rights reserved Manufacturer and Distributor STORZ BICKEL GMBH CO KG In Gruben cker 5 9 78532 Tuttlingen Germany Phone 49 7461 969707 0 Fax 49 7461 969...

Отзывы: