127
en
fr
es
it
nl
de
10. H OT A IR G EN ER ATO R
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – Guidance
Conducted RF
IEC 61000-4-6
Amplitude
modulated
Radio-frequency
electromagnetic
field
Amplitude
modulated
IEC 6100-4-3
3 V at 150 kHz-
80 MHz
6 V in ISM and
amateur radio
bands
80 % AM at 1
kHz
10 V/m at
80 MHz -
2.7 GHz
- Home
Healthcare
- Prof. Healtcare
- 80 % AM at 1
kHz
3 V at 150 kHz-
80 MHz
6 V in ISM and
amateur radio
bands
80 % AM at 1
kHz
10 V/m at
80 MHz -
2.7 GHz
- Home
Healthcare
- Prof. Healtcare
- 80 % AM at 1
kHz
WARNING: Portable and
mobile RF communica-
tions devices (including
their accessories such
as external antennas
and antenna cables)
should be used no closer
than 30 cm (12 inches)
to any part of the
VOLCANO MEDIC 2
Vaporization System,
including cables, speci-
fied by the manufacturer.
Otherwise, degradation
of the performance of
this equipment could
result.
WARNING: Portable and
mobile RF communica-
tions devices (including
their accessories, such
as external antennas
and antenna cables)
should be used no closer
than 30 cm (12 inches)
to any part of the
VOLCANO MEDIC 2
Vaporization System,
including cables, speci-
fied by the manufacturer.
Otherwise, degradation
of the performance of
this equipment could
result.
Содержание 01 01 M
Страница 2: ......
Страница 69: ...69 de en fr es it nl ...