Sealing Head Operation
S
EAL
S
PACING
Seals that progress along a
close-ended
length of tubing, as in a segmenting process,
must be spaced
no less
than the blood bag manufacturers recommen- dation, to
avoid rupture of the tubing due to pressure build-up.
Seals that progress along an open-ended length of tubing may be spaced as
desired, but ½ inch (1 cm) is the recommended minimum.
Multiple seals in close proximity are not required, nor are they recommended
when using the Power Source. If multiple seals must be used, they should be
spaced at least one inch (2.5 cm) apart, as shown in Figure 3-4, Multiple Seal
Figure 3-4, Multiple Seal Spacing
Warning: Never attempt to make segmentation seals closer than recommended
without validating for seal integrity.