b. Ceiling Mount: 

 The  ceiling  members  must  be  able  to 

support 100 pounds.

i.  The three chains, or cables, used must be rated at 

100 pounds minimum.  They must also be an equal 

distance from the ground to allow the 

Fume Hood

 to 

hang level to the ground.  The frame members used 

must also be able to support 100 hundred pounds.

ii.  Thread  the  Eye  Bolts  into  the  top  of  the  unit  a 

minimum of five turns. Tighten down the lock-nut 

with a washer in between the case and nut.

iii.  The unit can now be hung from the hanging devices 

chosen.

Note:

 Check local laws and regulations. If unable to install safely 

or if required by regulations, hire a certified technician to install 

the product.

5.  Once your Fume Hood has been securely mounted in your 

selected  location,  plug  one  end  of  the  Power  Cord  into  the  socket  labeled 

“Power” on the 

top of the unit, and then plug the other end into any standard grounded wall outlet. To test the unit, 

flip the 

On/Off/Remote

 swtich on the top of the unit into the 

On

 position and turn the selector switch 

on the bottom of the unit to either low, medium or high.

6.  If you are using Vaniman’s Remote Switch accessory, 

P/N VMC-A450

, test it with the Fume Hood by 

plugging it into the socket labeled Remote on the control panel on the top of the unit. Next, flip the 

On/Off/Remote

 switch on that same control panel to the Remote position. Ensure that the selector 

switch is either on low, medium, or high, then flip the switch on the Remote box into the 

On

 position. 

You should hear the Fume Hood turn on and moving air. Make sure to turn the Remote switch off 

when Fume Hood is not in use.

APPLICATION INSTRUCTIONS

1. 

The

 Fume Hood

  works  well  over  labratory  furnaces  or  ovens  that  produces  smoke  or  fumes. A 

minimum distance of four inches is required between the heat source and the bottom of the case (not 

including the skirt). In most cases, the closer the Fume Hood is to this distance from the source, the 

better the results as the fumes are kept from escaping. Some hotter furnaces may require a greater 

distance.  The bottom of the case temperature cannot exceed 

120 degrees Fahrenheit

.

2. 

The 

Fume Hood 

also works well for fumes from chemicles such as acrylic/monomer. If this is the 

desired application, Vaniman has a

 Tabletop

 

Stand, P/N 10516,

 and a 

Stand Window, P/N 10517,

 

to prevent fumes from reaching the technician’s face.

Optional Screw

Fig B

Содержание VMC-A450

Страница 1: ...Hood be sure that the location allows the power cord to be accessible at all times Additionally since the Pure Breeze Fume Hood draws air into the unit from the bottom and exhausts clean air from the...

Страница 2: ...2 Laying the Fume Hood flat with the control panel facing up unlatch the door from the left side and open Remove all filters from the unit and remove the plastic bags covering each filter The plastic...

Страница 3: ...each side for proper exhausting ventilation and access to the door latches i One piece of the bracket is mounted on the back of the unit at the factory ii Mount the second bracket piece with the open...

Страница 4: ...he selector switch on the bottom of the unit to either low medium or high 6 If you are using Vaniman s Remote Switch accessory P N VMC A450 test it with the Fume Hood by plugging it into the socket la...

Страница 5: ...filters and replace them in the opposite order then they were removed Order P N 97021 b The Secondary Filter is identified by its white media and is the second filter from the bottom in the stack This...

Страница 6: ...the On Off switch into the Off position unplug the power cord from the unit correct the problem then push the breaker reset button until it clicks 2 Thermal Protection The motor in the unit is protect...

Страница 7: ...vice disconnecting means cannot be locked securely fasten a prominent warning device such as a tag to service the panel c When servicing or replacing components unplug the power cord to ensure that th...

Отзывы: