1.

Remove the glass stage insert and unplug the main power cable and reflected
illuminator power cable from the rear of the microscope stand. Carefully lay
the microscope onto it’s back.

2.

Open the lamp access compartment located on the bottom of the microsope
base by turning the set knob (figure 57) counter-clockwise until the door
comes loose.

3.

When replacing, insert the new lamp into the lamp socket (figure 58). Make
sure that the pins on the lamp line up with the holes on the lamp socket. The
pins should slide freely into the holes with only slight resistance - do not
force.

4.

Close the compartment door and retighten the set knob. Follow the
instructions in the Transmitted Illumination Setup section to properly center
the lamp.

Replacing the Transmitted Illuminator Lamp

Before attempting to replace or remove the lamp,
UNPLUG THE MICROSCOPE FROM ANY POWER
SOURCE and allow to cool.

!

Be careful not to touch the glass bulb when replacing -
use a tissue or other medium to grasp the lamp. This
will prevent the oils in your hand from reducing lamp
life.  If contact is made with the glass bulb, clean with
rubbing alcohol and allow a brief drying period.

!

Replacement Lamp

 - 30W 12V Halogen (Cat. No. 1400-30W12VHL)

1.

Open the lamp housing door located at the rear of the reflected illuminator
box by pulling downward (figure 59).

3.

When replacing, insert the new lamp into the lamp socket.  Make sure that
the pins on the lamp line up with the holes on the lamp socket. The pins
should slide freely into the holes with only slight resistance - do not force.

4.

Close the compartment door and follow the instructions in the Reflected
Illumination Setup section to properly center the lamp.

Replacing the Reflected Illuminator Lamp

Replacement Lamp

 - 50W 12V Halogen (Cat. No. 1400-50W12VHL)

Figure 57

Figure 59

Contact us for current pricing and part numbers

Camera Kits

USB with Image Capture Software
DSLR
CCD Video
Video Eyepiece

Eyepiece Reticles

5mm Scale
10mm Scale
0.500” Scale
10mm Grid
Crosshair
Pointer
Metric & SAE Scale Calibration Plates

Objectives

50X LWD 0.70NA, 3.68mm Working Distance
60X LWD 0.75NA, 3.18mm Working Distance
80X LWD 0.80NA, 1.25mm Working Distance
100X LWD 0.85NA, 0.40mm Working Distance

Miscellaneous

Microscope Optical Cleaning Kit
4”x6” Lens Paper
Optical Tissue

Included Parts

Optional Accessories

Trinocular Head Assembly

1 ea.

Reflected Illuminator Assembly

1 ea.

Reflected Illuminator Lamp Housing

1 ea.

Stand

1 ea.

Brightfield Abbe Condenser

1 ea.

10X High Eyepoint Ultra-Widefield Eyepiece 2 ea.
5X Plan Infinity LWD Objective

1 ea.

10X Plan Infinity LWD Objective

1 ea.

20X Plan Infinity LWD Objective

1 ea.

40X Plan Infinity LWD Objective

1 ea.

Glass Stage Insert

1 ea.

Dispersion Filter

1 ea.

Blue Filter

1 ea.

Spare 30W/12V Halogen Lamp for
Transmitted Illuminator (#1400-30W12VHL) 1 ea.
Spare 50W/12V Halogen Lamp for
Reflected Illuminator (#1400-50W12VHL)

1 ea.

Spare 3A Fuse (#1400-FS5)

1 ea.

Power Cord

1 ea.

Rubber Eyecups

2 ea.

2mm Hex Wrench

1 ea.

3mm Hex Wrench

2 ea.

Operation Manual

1 ea.

Warranty Card

1 ea.

Dust Cover

1 ea.

For information about parts, accessories, or service -- contact your dealer directly
or contact VanGuard Microscopes at 1-800-423-8842.

Figure 58

set knob

lamp socket

lamp

socket

housing

door

For information about parts, accessories, or service -- contact your dealer directly
or contact VanGuard Microscopes at 1-800-423-8842.

Содержание 1442MMi

Страница 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Страница 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Страница 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Страница 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Страница 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Страница 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Страница 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Страница 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Страница 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Страница 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Страница 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Страница 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Страница 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Страница 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Отзывы: