background image

Components

Components

Components

Components

Components

7

Maintenance

Maintenance

Maintenance

Maintenance

Maintenance

22

1.

Remove the fuse holder from the rear of the microscope by using a slotted
screwdriver to push downward on the fuse comparment door slot while
prying outward (figure 60).

2.

Remove the fuse from the fuse holder (figure 61).

3.

If the fuse is blown replace by inserting a new fuse into the fuse holder.

4.

Push the fuse holder back into the fuse holder slot until fully seated.

Replacing the Fuse

Before attempting to replace or remove the fuse,
UNPLUG THE MICROSCOPE FROM ANY POWER
SOURCE.

!

Replacement Fuse 

- 3A 250V (Cat. No. 1400-FS5)

If the microscope is plugged in but the illuminator is not turning on, the fuse
could be blown. The instructions below detail how to check for a blown fuse.
You can usually tell a blown fuse by if the wire inside is broken, or the glass is
blackened. Sometimes there will be no visible indication though and it will be
necessary to check the fuse with a conductivity meter or to replace it with a
new fuse.

For information about parts, accessories, or service -- contact your dealer directly
or contact VanGuard Microscopes at 1-800-423-8842.

Figure 60

Figure 61

Colored Filter

Wheel

Field Diaphragm

Lamp Housing

Set Screw

For information about parts, accessories, or service -- contact your dealer directly
or contact VanGuard Microscopes at 1-800-423-8842.

Aperture Diaphragm

Содержание 1442MMi

Страница 1: ...Printed in the USA VGMNL062512 1400MMi Operation Manual Model 1442MMi www veegee com 800 423 8842 sales veegee com support veegee com...

Страница 2: ...croscope system Modifying the product in any manner for use not originally intended shall automatically void the manufacturer s warranty In no event shall VEE GEE Scientific Inc be held liable for any...

Страница 3: ...surfaces are made Shock Hazard failure to follow instructions could lead to electrical shock Safety Precautions Before replacing the lamp or fuses always turn off the power switch and disconnect the p...

Страница 4: ...ymptom Image through eyepieces is too dim Step Possible Cause HowToTest TestResult Solution For More Info 1 2 Variable lighting control VLC is turned down too far Visually inspect VLC is turned up VLC...

Страница 5: ...turned down Go to next step Turn VLC up Page16 Microscope not getting power Inspect outlet Good outlet Outlet not good Go to next step Plug power cord into working outlet Blown fuse Visually inspect...

Страница 6: ...an to prevent contaminate ingress is essential to the longevity of the objectives All other parts can be cleaned with a paper towel and mild detergent It s not recommended to use rubbing alcohol for c...

Страница 7: ...eplacement Fuse 3A 250V Cat No 1400 FS5 If the microscope is plugged in but the illuminator is not turning on the fuse could be blown The instructions below detail how to check for a blown fuse You ca...

Страница 8: ...resistance do not force 4 Close the compartment door and follow the instructions in the Reflected Illumination Setup section to properly center the lamp Replacing the Reflected Illuminator Lamp Replac...

Страница 9: ...lyzer knob located on the right side of the reflected illuminator must be pushed inward to engage this function figure 55 1 Carefully remove all components from foam packaging and plastic bags Use the...

Страница 10: ...ve In addition you will need to fully open the aperture iris in order to prevent vignetting and leave the condenser height the same as when originally setup using the 10X objective 5 To control contra...

Страница 11: ...er installed but should you need to reinstall at any point follow the directions below 13 Rotate the nosepiece so that either the 5X objective or an empty slot is in the light path 14 Using the coarse...

Страница 12: ...making adjustments to the illuminator housing or lamp fixture 3 Ensure that the illumination selector knob is pushed all the way inward R figure 17 Figure 15 Figure 17 Figure 18 Figure 19 4 Rotate the...

Страница 13: ...b adjusts for latitudinal X axis and the lower for longitudinal Y axis movement figure 21 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 10 Rotate each of these in succession until the field of vie...

Страница 14: ...and should be checked and readjusted before each use Focus Adjustment 18 Focusing on your specimen is achieved through rotating the coarse and or fine focusing knobs These are located on either side o...

Отзывы: