REV 2
CT-7500 S2 USER’S MANUAL
4
Performing Diagnostics
Follow the steps below to perform diagnostics and confirm that the adapter is calibrated
properly:
a.
Move the resistor transducer to one end of its travel.
b.
From the “RUNNING” screen, press either the
[
▲
]
or
[
▼
]
key to access the Main Menu.
The following screen will be displayed:
The DIAGNOSTIC option should be highlighted. Press the
[ENTER]
key.
c.
The following screen will be displayed:
Move the resistor transducer and observe the distance values on the screen:
Move the resistor transducer to the other end of its travel and observe the screen:
The value should be the distance between the two endpoints of the transducer. If this
value is not correct, you should calibrate the adapter.
ENCODER TRAVEL:
5.00 IN 12.7 CM
“ENTER” WHEN DONE
ENCODER TRAVEL:
2.23 IN 5.66 CM
“ENTER” WHEN DONE
ENCODER TRAVEL:
0.00 IN 0.00 CM
“ENTER” WHEN DONE
CALIBRATE ENCODER
ADJUST CONTRAST
<DIAGNOSTIC>