Calibration
1. Should calibration be necessary, all that is required is a 25 lb precision weight (do NOT
use bags of flour or other packaged goods for weights - contact the Lift Manufacturer to
purchase precision weights).
2. To enter the calibration mode, first press and hold the “ON/ZERO” pad. While holding
the “ON/ZERO” pad press and hold the “LB/KG” pad. After approximately 3 seconds the
message on the display will change from “Zero” to CAL”. Release both pads at this time
and the display will show the “C 0” message.
3. Make sure the sling is empty and press the “ON/ZERO” pad. The indicator will count
down from -
8
- to -0- while taking readings and store the zero point. When finished,
the indicator will show the “C25” message.
4. Place a 25 lb test weight in the sling and press the “ON/ZERO” pad once again. The
indicator will count down from -
8
- to -0- while taking readings and calculate and store
the span calibration. The indicator then automatically returns to the normal weighing
mode. Calibration is now complete.
Safety Alert!
It has been brought to our attention that the hanging VANDERSCALE to be installed on your
VANDER-LIFT II
TM
can become lodged on the boom, thereby causing inaccurate weights.
The potential also exists that when so lodged it may break the safety pin, thereby causing
the scale to unhook.
THIS COULD CAUSE AN INJURY TO A RESIDENT AND/OR STAFF
MEMBER.
Shown below is a part called a bushing. This part is designed to help prevent this problem
from occurring.
Installation of Bushing / IMS & SR Scales
FLAT HEAD SCREW (Steps 1 & 3)
BUSHING (Step 2)
(Patent Pending)
Step 1: Use a standard screwdriver and remove flat head screw.
(See above)
Step 2: Slide Bushing into place as shown above.
Step 3: Replace flat head screw and tighten firmly.
39
November 2017
For your nearest Distributor call, 1-800-694-4525
39