Facility:
Contract Dates:
To:
Address:
Scheduled:
Actual:
Room Number:
Next PM Check:
Lift Model:
Number of Lifts Since Last PM:
Lift Serial Number:
Additional Service Needed?
YES
NO
Lift Functions:
Description:
UP, DOWN, EMRG. DOWN,
TRAVERSE, ON/OFF, EMRG.
ON/OFF
Check functions using the
buttons on the lift.
YES
NO
N/A
YES
NO
Mechanical Functions:
Description:
Load Test Per ISO10535
Lift 100% of the load capacity
of the Motor 20" off of the
ground
YES
NO
N/A
YES
NO
Trolley Wheels
Inspect wheels for flat spots,
and excessive wear and tear
YES
NO
N/A
YES
NO
Motor, Gears, and
Traversing Drive
Inspect for damage and
excessive noise
YES
NO
N/A
YES
NO
Hange
r
Bar
Inspect for damage
YES
NO
N/A
YES
NO
Fraying of Strap Edges
Lower the strap down to the
ground and inspect the full
length of the edges.
YES
NO
N/A
YES
NO
Pneumatic Hand Control:
Description:
UP, DOWN, EMRG. DOWN,
TRAVERSE, ON/OFF
Press each button on hand
control for 10 seconds, make
sure button function works
continuously.
YES
NO
N/A
YES
NO
Airline Tubing
Inspect for damage/leaks
YES
NO
N/A
YES
NO
Grommet Connectors
Check that they are tight
YES
NO
N/A
YES
NO
Limit Switches:
Description:
Hold
UP
until the carry bar is at
the top. Motor should stop
automatically.
YES
NO
N/A
YES
NO
Let the strap out all the way to
the ground. Motor should stop
before strap winds backwards.
YES
NO
N/A
YES
NO
While using either
UP
or
Down
,
angle the strap more than 15
degrees. Motor should stop
automatically.
YES
NO
N/A
YES
NO
Slack Tape Switch
Hold
DOWN
and lift up on the
carry bar. Motor should stop
automatically.
YES
NO
N/A
YES
NO
Electrical Functions:
Description:
Battery Voltage:
Check batteries with multimeter
and confirm 12V reading while
engaging the motor during load
test.
YES
NO
N/A
YES
NO
Charger:
Engage unit with charger and
check that unit is charging
properly. Light on the lift should
turn amber.
YES
NO
N/A
YES
NO
Miscellaneous
Description:
LCD Screen:
Check that LCD works properly
YES
NO
N/A
YES
NO
LED Light:
Check that light turns green
when on, dark when off, and
amber when charging.
YES
NO
N/A
YES
NO
Casing
Check for cracks or wear in case
YES
NO
N/A
YES
NO
PM Reset
Use hand control to
enter
program mode. Log # of lifts in
upper right corner on this
sheet. Reset the PM Counter.
YES
NO
N/A
YES
NO
Vancare, Inc.:
Sign
Date
Customer:
Sign
Date
Upper Limit Switch
Pass Inspection:
Fixed:
Problems:
Problems:
Fixed:
Pass Inspection:
Fixed:
Problems:
Problems:
Problems:
Pass Inspection:
Problems:
Fixed:
Ceiling Lift - Preventative Maintenance
Pass Inspection:
Pass Inspection:
Pass Inspection:
Fixed:
Fixed:
P-600 Technical Manual
REV: 101316
Page 34
Содержание P-600
Страница 1: ...Manufactured for P 600 Technical Manual ...
Страница 10: ...P 600 Technical Manual REV 101316 Page 10 ...
Страница 12: ...P 600 Technical Manual REV 101316 Page 12 ...
Страница 13: ...P 600 Technical Manual REV 101316 Page 13 ...
Страница 23: ...P 600 Technical Manual REV 101316 Page 23 ...
Страница 24: ...Service Parts List P 600 P 600 Technical Manual REV 101316 Page 24 ...
Страница 25: ...P 600 Technical Manual REV 101316 Page 25 ...
Страница 26: ...P 600 Technical Manual REV 101316 Page 26 ...
Страница 27: ...P 600 Technical Manual REV 101316 Page 27 ...
Страница 28: ...P 600 Technical Manual REV 101316 Page 28 ...
Страница 29: ...P 600 Technical Manual REV 101316 Page 29 ...
Страница 30: ...P 600 Technical Manual REV 101316 Page 30 ...
Страница 31: ...P 600 Technical Manual REV 101316 Page 31 ...
Страница 32: ...P 600 Technical Manual REV 101316 Page 32 ...