Monopolar Instrument Receptacles
Controls, Indicators,
and Receptacles
Force FX-8C Service Manual
2-7
Monopolar Instrument Receptacles
You can connect a footswitching or handswitching monopolar instrument
to the monopolar receptacles. Some footswitching instruments may
require a single-pin adapter (E0502 Series) or E0017, available from
Valleylab.
Connect one monopolar instrument to the Monopolar 1/CEM Instrument
receptacle:
• A single-pin footswitching instrument or a three-pin handswitching
instrument
or
• A four-pin CUSA handpiece with CEM nosecone. (The CEM indicator
in the upper right of the front panel illuminates green. Refer
to Section 4, Connecting the CUSA Handpiece with CEM Nosecone, in the
Force FX-8C Electrosurgical Generator User’s Guide
.
Connect one monopolar instrument to the Monopolar 2 Instrument
receptacle:
• A single-pin footswitching instrument or a three-pin handswitching
instrument
REM Alarm Indicator
This indicator illuminates red until you properly apply a REM patient
return electrode to the patient and connect it to the generator. Then the
indicator illuminates green. (When you connect an electrode without the
REM safety feature, the indicator does not illuminate.)
If the REM system senses an alarm condition, the indicator flashes red
until you correct the alarm condition—then the indicator illuminates
green. (If you are using a return electrode without the REM safety feature,
the red indicator light is extinguished when you correct the alarm
condition.)
Warning
The instrument receptacles on this generator are designed to accept only one
instrument at a time. Do not attempt to connect more than one instrument at a
time into a given receptacle. Doing so will cause simultaneous activation of the
instruments.
Содержание Force FX-8C
Страница 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Страница 30: ...Notes 2 12 Force FX 8C Service Manual...
Страница 56: ...Notes 3 26 Force FX 8C Service Manual...
Страница 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Страница 84: ...Notes 4 28 Force FX 8C Service Manual...
Страница 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...
Страница 190: ...Generator Assembly 9 2 Force FX 8C Service Manual Generator Assembly Figure 9 1 Generator assembly...
Страница 191: ...Generator Assembly Service Parts Force FX 8C Service Manual 9 3 Figure 9 2 Generator assembly continued...
Страница 194: ...Front Panel Assembly 9 6 Force FX 8C Service Manual Front Panel Assembly Figure 9 3 Front panel assembly...
Страница 195: ...Front Panel Assembly Service Parts Force FX 8C Service Manual 9 7 Figure 9 4 Front panel assembly continued...
Страница 216: ...Notes 9 28 Force FX 8C Service Manual...