Performance Characteristics
3-2
Force FX-8C Service Manual
Dimensions and Weight
Operating Parameters
Transport and Storage
Width:
35.6 cm (14 in.)
Depth
45.7 cm (18 in.)
Height
11.1 cm (4 3/8 in.)
Weight
< 8.2 kg (< 18 lbs.)
Ambient temperature
range
10° to 40°
C (50° to 104° F)
Relative humidity
30% to 75%, noncondensing
Atmospheric pressure
700 to 1060 millibars
Warm-up time
If transported or stored at temperatures outside the
operating temperature range, allow one hour for the
generator to reach room temperature before use.
Ambient temperature
range
-40° to 70° C (-40° to 158° F)
Relative humidity
10% to 100%, condensing
Atmospheric pressure
500 to 1060 millibars
Duration of storage
If stored longer than one year, the battery must be replaced
and a full checkout, including calibration, must be
completed before use. For instructions, refer to Section 5,
Calibrating the Generator, in this manual.
Содержание Force FX-8C
Страница 1: ...Service Manual Force FX 8C Electrosurgical Generator with Instant Response Technology...
Страница 30: ...Notes 2 12 Force FX 8C Service Manual...
Страница 56: ...Notes 3 26 Force FX 8C Service Manual...
Страница 58: ...Block Diagram 4 2 Force FX 8C Service Manual Block Diagram...
Страница 84: ...Notes 4 28 Force FX 8C Service Manual...
Страница 156: ...Interconnect Diagram 7 2 Force FX 8C Service Manual Interconnect Diagram...
Страница 190: ...Generator Assembly 9 2 Force FX 8C Service Manual Generator Assembly Figure 9 1 Generator assembly...
Страница 191: ...Generator Assembly Service Parts Force FX 8C Service Manual 9 3 Figure 9 2 Generator assembly continued...
Страница 194: ...Front Panel Assembly 9 6 Force FX 8C Service Manual Front Panel Assembly Figure 9 3 Front panel assembly...
Страница 195: ...Front Panel Assembly Service Parts Force FX 8C Service Manual 9 7 Figure 9 4 Front panel assembly continued...
Страница 216: ...Notes 9 28 Force FX 8C Service Manual...