No.Q-0600-0057-05

8. Maintenance, Inspection, and Adjustment

It is recommended to perform periodical inspection at least once a 
year although the appropriate inspection interval depends on the 
operating environment.
In addition, check the zero point every 3 months, and perform zero 
adjustment when necessary before use.
Normally, this unit requires only zero adjustment for maintaining 
high accuracy.    When this unit needs recalibration, send it to 
VALCOM or perform calibration by following the procedure below.

1) Prepare a calibrator with guaranteed accuracy.
2) If the input must correspond to the "0" indication, make sure that 

the indicated value is "0".

3) Make sure that the reading falls within the limits when the rated 

pressure is applied and when 1/2 of the rated pressure is 
applied.

4) If the reading is out of the limits, perform calibration by following 

the procedure below.

1. Remove the outer frame of the front panel, and then take the 

front panel out by hooking the notch with a screwdriver, etc.   
(When the front panel is secured with the panel mounting 
bracket, loosen it first.)

2. If the input must correspond to the "0" indication, turn the 

indication zero adjustment trimmer to set the indicated value to 
"0".
(For absolute pressure, zero the reading when 20Pa

abs or 

lower is applied.)

3. While applying the rated pressure, adjust the reading by 

turning the reading span-adjusting control.

4. Perform the above-mentioned calibration again for 

confirmation, and then check the non-linearity, etc. under 1/2 
of the rated pressure.

5.  In  addition,  adjust  the  analog  output  by  following  the 

above-mentioned procedure.

9

Содержание VPRHS-U

Страница 1: ...uy your excess underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing R...

Страница 2: ...Keys 3 Terminal Connections 4 Particularly Important Precautions 5 Zero shifting Function 6 Optional Peak holding Function 7 Optional Auto zeroing Function 8 Maintenance Inspection and Adjustment 9 I...

Страница 3: ...specified power supply voltage will cause fire shock hazards or failures Never modify the sensor or its accessories Neglecting this will cause fire shock hazards or failures Never touch any terminal o...

Страница 4: ...injury Be sure to confirm that power is turned OFF before relocating the sensor or changing the wiring Neglecting this may cause shock hazards or failures When one of abnormal conditions shown below i...

Страница 5: ...No Q 0600 0057 05 1 General Specifications 4...

Страница 6: ...ommon terminal for control input 4 Sensor connection D INPUT 13 SET1 output S N O SO Unused 5 Sensor connection E Shield Analog output 14 SET1 output S N C SO N O 6 Analog output 15 SET1 output COM 7...

Страница 7: ...g the sensor cable in parallel with a power line etc or in the same conduit Securely ground the grounding terminal F G frame ground 4 Particularly Important Precautions Be sure to use the amplifier an...

Страница 8: ...measurement is deviated Remove the outer frame of the front panel and then take the front panel out by hooking the notch with a screwdriver etc Place the sensor in a state where the reading should be...

Страница 9: ...easurement operation P Peak holding 7 Optional Auto zeroing Function Short circuiting the terminal No 11 Z and the terminal No 12 COM automatically zeros the current reading When the terminals are ope...

Страница 10: ...when the rated pressure is applied and when 1 2 of the rated pressure is applied 4 If the reading is out of the limits perform calibration by following the procedure below 1 Remove the outer frame of...

Страница 11: ...ics Analog input schematic A2 A3 A4 and A5 Control input schematic P and Z Analog output schematic A1 A2 A4 and A5 To internal CPU board 24V Control input 10K COM 0 1uF DG Output AG 100 1 4W Internal...

Страница 12: ...11 VPRVS Series VPRHS Series 48 9 9 9 1 9 9 9 96 2 1 1 5 5 104 89 5 41 5 17 5 127 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 9 9 9 48 SET1 1 96 9 9 9 SET2 2 5 5 104 89 5 8 7 6 5 4 3 2 17 16 15 14 1...

Страница 13: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...

Отзывы: