No.Q-0600-0057-05
* When you must perform calibration, follow the procedure
described in "8. Maintenance, Inspection and Adjustment."
(*While performing calibration, do not touch any control which
does not need adjustment.)
* When a measurable limit (FS.
±
approx. 2%) is exceeded, the
display starts flashing.
(The reading remains unchanged.)
When zero-shifting has been performed, the zero-shifting amount
is added to or subtracted from the original measurable limits.
5. Zero-shifting Function
Zero adjustment can be performed by operating the
zero-shifting key hidden behind the front panel.
Perform zero adjustment when, for example, the zero point
for measurement is deviated.
* Remove the outer frame of the front panel, and then take
the front panel out by hooking the notch with a screwdriver,
etc.
* Place the sensor in a state where the reading should be
zero.
* Pressing the zero-shifting key for at least 2 seconds, the
reading is forcefully zeroed. (Zero shifting)
* Pressing the zero-shifting key for 5 seconds or longer,
zero shifting is canceled.
* Do not operate any switch other than the zero-shifting key.
* When the difference between the readings before and after
zero shifting becomes
±
2%FS or more, a warning
message " " is temporarily shown.
(Z e ro sh if t in g is p e rf orm e d eve n if th e w a rn in g m e ssa ge " "
is s h ow n . )
(I f t h e wa rn in g m e ss a ge " " is s h o wn , it m ay b e im p o ss ib le
to m e a su re F S .)
(W h e n F S n e e d s t o b e m ea s u re d , ca n c e l ze ro s h if t in g, a n d m a ke
a d ju st m e n t w it h th e re ad in g ze ro -ad ju st in g c on t ro l. )
-
7
-