Chapter 6 ________________________________________________________ Connection Options
VAISALA_______________________________________________________________________ 79
Setting Fields
NOTE
You can add the Id information field in the supervisor data message to
provide identifying information in addition to the transmitter address.
See
Supervisor Message on page 144
. The information field is set as
part of the factory settings. See
General Unit Settings on page 204
.
You can only modify it with the Vaisala Configuration Tool.
a
=
Device address
XU
=
Device settings command in ASCII and NMEA 0183
XXU
=
Device settings command in SDI-12
[A]
=
Address: 0 (default) ... 9, A ... Z, a ... z
[M]
=
Communication protocol:
A = ASCII, automatic
a = ASCII, automatic with CRC
P = ASCII, polled
p = ASCII, polled, with CRC
N = NMEA 0183 v3.0, automatic
Q = NMEA 0183 v3.0, query (= polled)
S = SDI-12 v1.3
R = SDI-12 v1.3 continuous measurement
[T]
=
Test parameter (for testing use only)
[C]
=
Serial interface:
1 = SDI-12
2 = RS-232
3 = RS-485
4 = RS-422
[I]
=
Automatic repeat interval for Composite data
message:1 ... 3600 s, 0 = no automatic repeat
[B]
=
Baud rate: 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200
[D]
=
Data bits: 7/8
[P]
=
Parity:
O = Odd
E = Even
N = None
[S]
=
Stop bits: 1/2