-A
(ON) to be able to enter new linear and multipoint corrections to the
transmitter.
ON
OFF
SW1
SW2
SW3
SW4
SW5
SW6
SW7
SW8
S1
F
IGURE
6-1
Switch SW4 in write ENABLE position (ON)
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).
CORR Listing linear and multipoint pressure corrections
CORR <cr>
The CORR command is used to indicate the status of linear and
multipoint pressure corrections. In case the status is ON, the
corrections and valid date of calibration are listed. The listing varies
according to the number of transducers in the transmitter. The
transmitter lists * in the second column if there exists only one
pressure transducer in the transmitter.
Example of the listing of a transmitter with one pressure transducer:
>corr
Linear adjustments
ON
Reading Correction
Reading Correction
500.010
0.120*******.**********.***
1100.320
0.150*******.**********.***
Multipoint adjustments
ON
Reading Correction
Reading Correction
499.660
-0.110*******.**********.***
599.110
-0.080*******.**********.***
698.580
-0.060*******.**********.***
800.950
-0.030*******.**********.***
900.400
0.010*******.**********.***
947.200
0.020*******.**********.***
999.840
0.050*******.**********.***
1099.090
0.070*******.**********.***
All manuals and user guides at all-guides.com
all-guides.com