Example of setting the output format:
>form
P1 " " UUU " " T " " UU " " RH " " UUU \r \n
? 4.1 P " " UUU " " "Tend" " " 1.0 A " " 3.1 T UU " "
3.0 RH UUU \r \n
>send
1007.9 hPa Tend 7
24.9'C
28%RH
>
Inspecting the output format:
>form
P1 " " UUU " " T " " UU " " RH " " UUU \r \n
?<esc>
Use E
SC
-key to abort without changing the settings.
EFORM Defining the error output format
EFORM <cr>
The EFORM command can be used to define a user specific error
output format for the serial line. In case of an error, the transmitter
outputs the defined format instead of *****.
Example of an EFORM definition:
>eform <cr>
? "ERROR" \r \n <cr>
>send
1007.8 hPa
OK
(correct operation)
>send <cr>
ERROR
(incorrect operation)
>
Any previous EFORM definition may be removed with the following
command:
>eform * <cr>
>