SEND mode
In the SEND mode, the transmitter will automatically output one
message at power-up or when triggering the transmitter on using pin 4
of the transmitter. The trigger signal must be a TTL level signal. TTL
HIGH (5 VDC) turns the transmitter ON and TTL LOW (0 VDC)
turns the transmitter OFF.
Examples of outputting with the prompt (ECHO ON) and without the
prompt (ECHO OFF):
>echo <cr>
Echo
:
ON
>smode send <cr>
Serial mode
: SEND
>reset <cr>
1013.25 hPa
>
>echo off <cr>
Echo
:
OFF
reset <cr>
(text invisible)
1013.25 hPa
Example of outputting a minimum amount of (six) characters by
excluding the <cr> and <lf> characters, the pressure unit and the
decimal point with the FORM command:
>form <cr>
4.2 P " " UUUU \r \n
? 6.0 P <cr>
>unit Pa
P unit
: Pa
>smode send <cr>
Serial mode
: SEND
>echo off <cr>
Echo
:
OFF
reset <cr>
(text invisible
)
101325
or
(power-up)
101325