Recommended to copy dry road coefficient values from one sensor to another when dry road
reference has been done on a similar road surface before and surface coefficients are known.
1.
Check the values from a sensor that you are using in the same area. Use the following message:
GET PARAMETER
The requested parameters are
0x53
,
0x54
, and
0x55
.
2.
For the sensor that you want to calibrate, give the values for the 3 parameters (Road coefficient,
laser 1, laser 2, and laser 3). Use the following message:
SET ROAD COEFFICIENTS
Setup Guide
M212169EN-D
23
Option C: copying coefficient values from another sensor
The sensor to which you are copying the values:
•
Must have plate reference done
•
Must be used in the same area as the sensor from which you copy the values