Calibration step 2: plate reference
The reference setting is performed using the reference plate.
Requirement
Description
Mobile sensor placement
Installed on vehicle
Vehicle position
At a standstill
Reference plate surface
Dry, no condensation
Conditions, indoors
Clean and dry space
Conditions, outdoors
No precipitation or fog
Plate adjustment minimizes the effects caused by different installation heights and differences
between individual mobile sensors. Repeat this procedure when changing the installation position.
Keep the reference plate container closed when not in use. Handle the reference plate
with care and keep it clean. Avoid scratching the surface. The reference plate is made of optically
reflective material.
CAUTION!
1.
Check that the markings in the body and hood are aligned.
2.
Open the reference plate container. Check that the reference plate is clean and dry. If it is not,
carefully remove the dirt.
3.
Place the reference plate as shown.