4.
Check that the mobile sensor is powered and operational (green LED in the mobile sensor is
blinking).
5.
Check that the sensor is not reporting errors.
Use the
SEND DATA
request, and check the unit status information and error bits from the
SEND DATA
response.
SEND DATA
6.
Give the following command and parameter:
SET REFERENCES
PLATE
This changes reference value parameters
0x50
,
0x51
, and
0x52
(1 parameter for each laser).
•
When started, changes status bit 1 (reference setting ongoing) to
1
= ongoing.
•
When finished, changes back to
0
= not ongoing.
The adjustment takes typically 30 seconds.
7.
Check that plate adjustment was successful.
Use the
SEND DATA
request, and check the unit status information from the
SEND DATA
response.
SEND DATA
The command was successful if status bits 10 - 13 (
reference setting interrupted due
to...
) all have value
0
(no error).
If the adjustment is not successful, repeat the procedure.
Setup Guide
M212169EN-D