12. Start the high point output test by selecting
Output high point > OK
.
The analog channel starts to send the high point test output (10 V in this example).
13. After the high point test output has stabilized, check the multimeter reading. Let the test
stay in progress and press to return to the previous menu level, where you can enter
the reading to Indigo80.
Do not stop the test at this point.
14. After pressing , enter the reading of the multimeter to Indigo80:
a. Select
Measured value, low point
.
b. Use the arrow buttons to set the reading, and then select
Set
.
15. To save the corrections to the output level, scroll down and select
Activate adjustment >
OK
. Confirm by selecting
OK
.
The corrections are shown under the
Analog output 1 adjustment
view, in
Low point
offset
and
High point offset
.
16. To verify that the output level of the transmitter is now accurate, repeat the low point and
high point output tests and check that the multimeter reading matches the test output.
17. To enter the adjustment date and information about the adjustment, scroll down the
Analog output 1 adjustment
view and set
Calibration date
and
Calibration text
.
18. Make sure that the adjustment mode is disabled with the
Adjustment mode on/off
selection.
19. Exit the menu by pressing or return to the main menu by pressing .
20. Switch off the power supply, remove the multimeter wiring and replace the analog output
wiring.
21. After completing the wiring, close the transmitter cover and switch on the power supply.
More information
‣
Opening transmitter cover (page 16)
‣
Analog output terminals (page 23)
‣
Connecting to Indigo80 handheld indicator (page 36)