User's Guide ______________________________________________________________________
96 __________________________________________________________________ M210566EN-J
Stopping Continuous Outputting
Use the
S
command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
S
<cr>
See command
SMODE
to change the default (power-up) operation
mode.
Outputting Reading Once
Use the
SEND
command to output the reading once in STOP mode. The
output format depends on which parameters the transmitter can output.
SEND
<cr>
Examples:
>
send
RH= 98.4 %RH T= 31.1 'C
>
send
RH= 98.4 %RH T= 31.1 'C Td= 36.0 'C Tdf= 36.0 'C a= 42.4
g/m3 x= 38.8 g/kg Tw= 30.8 'C ppm= 62414 pw= 59.53 hPa
pws= 60.52 hPa h= 130.7 kJ/kg
Assign an Alias for the SEND Command
Use the
SCOM
command to assign a new command that works like the
SEND
command. The standard
SEND
command of the transmitter will
always function normally whatever the
SCOM
definition may be.
Command names are case-insensitive.
SCOM
<cr>
Example (assigning the command
MEASURE
as an alias for the
SEND
command):
>
scom
Send command : ?
measure
>
measure
RH= 23.8 %RH T= 21.4 'C
>
You can remove the
SCOM
definition by pressing
Esc
when entering the
command name:
>
scom
Send command : measure ?
<esc>
>