Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 117
Example (without data logger module):
>
dir
File description
Oldest data available
No. of points
1 RH (10 s intervals) 2008-04-11 23:41:10
135
2 RH (90 s intervals) 2008-04-11 20:41:11
135
3 RH (12 min intervals) 2008-04-10 21:03:41
135
4 RH (2 h intervals) 2008-03-31 18:03:41
135
5 RH (12 h intervals) 2008-02-04 12:03:41
135
6 RH (3 d intervals) 2007-03-04 00:03:41
135
7 T (10 s intervals) 2008-04-11 23:41:11
135
8 T (90 s intervals) 2008-04-11 20:41:11
135
9 T (12 min intervals) 2008-04-10 21:03:41
135
10 T (2 h intervals) 2008-03-31 18:03:41
135
11 T (12 h intervals) 2008-02-04 12:03:41
135
12 T (3 d intervals) 2007-03-04 00:03:41
135
13 Tdf (10 s intervals) 2008-04-11 23:41:11
135
14 Tdf (90 s intervals) 2008-04-11 20:41:11
135
15 Tdf (12 min intervals) 2008-04-10 21:03:41
135
16 Tdf (2 h intervals) 2008-03-31 18:03:41
135
17 Tdf (12 h intervals) 2008-02-04 12:03:41
135
18 Tdf (3 d intervals) 2007-03-04 00:03:41
135
>
PLAY
Use the
PLAY
command to output the selected file to the serial line. If
the data logger module is installed, you can specify an interval to be
outputted.
Data in the output is <TAB> delimited. This is compatible with most
spreadsheet programs. Before giving the command, set the local date and
time with
TIME
and
DATE
commands, if needed.
PLAY
[
x
] [
start_date start_time end_date end_time
]<cr>
where
x
= Number of the data file that will be outputted, range 0...21.
The numbers correspond to the output of the DIR command;
refer to the example on page 116.
Selecting number 0 will output all data files.
start_date = Starting date of the interval to be outputted. Must be given in
the following format: yyyy-mm-dd.
start_time = Starting time of the interval to be outputted. Must be given in
the format hh:mm:ss or h:mm.
end_date = Ending date of the interval to be outputted. Must be given in
the following format: yyyy-mm-dd.
end_time = Ending time of the interval to be outputted. Must be given in
the format hh:mm:ss or h:mm.