SCALING THE ANALOG OUTPUT VIA THE SERIAL BUS 

Serial communication settings 

The analog output scaling can be done using the serial bus and computer with suitable terminal software. 
Connect the DMT242 to a serial bus via the RS-232 interface by using the following settings: 

Baud rate: 

2400, Parity: none, Data bits: 8, Stop bits: 1

. The serial cable (DMT242RS) can be ordered from Vaisala.  

Scaling of the dewpoint output 

Scale the dewpoint parameter by giving the command 

ascl xx yy<ENTER>

 where xx = the low limit 

(

°

C or ºF) and yy = the high limit (

°

C or ºF). The factory setting for serial measurement unit is 

Centigrade (

°

C). To change the unit for Fahrenheit (ºF) use command 

unit xx <ENTER>

 where xx = 

n

 

for non-metric (ºF) and xx = 

m

 for metric (

°

C) units. 

ACCESSORIES 

Order Code 

Description 

HM47280 Stainless 

steel sintered filter 

DMT242RS 

RS-232 serial line cable for PC (with female D connector) 

210662 

1/2" NPT adapter 

 

GUARANTEE 

Vaisala issues a guarantee for the material and workmanship of this product under normal operating 
conditions for one (1) year from the date of delivery. Exceptional operating conditions, damage due to 
careless handling and misapplication will void the guarantee. 