SCALING THE ANALOG OUTPUT VIA THE SERIAL BUS
Serial communication settings
The analog output scaling can be done using the serial bus and computer with suitable terminal software.
Connect the DMT242 to a serial bus via the RS-232 interface by using the following settings:
Baud rate:
2400, Parity: none, Data bits: 8, Stop bits: 1
. The serial cable (DMT242RS) can be ordered from Vaisala.
Scaling of the dewpoint output
Scale the dewpoint parameter by giving the command
ascl xx yy<ENTER>
where xx = the low limit
(
°
C or ºF) and yy = the high limit (
°
C or ºF). The factory setting for serial measurement unit is
Centigrade (
°
C). To change the unit for Fahrenheit (ºF) use command
unit xx <ENTER>
where xx =
n
for non-metric (ºF) and xx =
m
for metric (
°
C) units.