U

SER

'

G

UIDE 

 

June

 

2008

 

|

 

M010036EN-D 

5.

 

Insert the wired screw terminal back into the connector excatly in the position indicated in Figure 3. 
Push the fixing screw through the connector. Fasten the cable clamp. Insert the wired connector into 
the counter connector. 

NOTE 

The connection is incompatible if the positioning of the screw terminal is NOT as 
indicated in Figure 3. Fasten the fixing screw. 

 
6.

 

Install the back cover allowing the cable to run through the hole in the cover. The transmitter is ready 
for use. 

NOTE 

When the power is switched on, wait about 7 minutes before taking measurements. 
Start-up self-diagnostics freeze the output during the first operation minutes. 

 

Figure 3 Wiring and Installation of Connector 

 
CALIBRATION AND MAINTENANCE 

Replacing the sintered filter 

If the sintered filter is dirty it can prolong the response time of the measurement. If the filter needs to be 
changed unscrew the filter and replace the old filter with a new one. 

Calibration 

It is recommended to perform a calibration check every other year. A field check can be carried out by 
using a calibrated reference probe and comparing the readings measured with the transmitter and the 
reference probe. The Vaisala DRYCAP

®

 Hand-Held Dewpoint Meter DM70 is ideal for confirming the 

performance of the transmitter in the field. By using a connection cable the readings of DMT242 and 
DM70 can be viewed simultaneously on the display of DM70. If there is need for adjustement contact 
Vaisala Instruments Service Centers or local Vaisala representative. 

 

SCREW TERMINAL MUST BE INSTALLED TO THE 
CONNECTOR IN THE POSITION OF THE FIGURE 

(

TERMINALS 

2

 AND 

3

 AS UPPER EDGE

1

 

=   Connector 

2

 

=   Screw terminal of the connector 

3

 

=   Connector 

4

 

=   Cable clamp 

5

 

=   RS-232 serial port 

6

 

=   Terminals 2 and 3 

7

 

=   Three-wire cable 

8

 

=   Pressure setting switches 

9

 

=   Fixing screw 