![Vaisala Drycap DMT152 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362063.webp)
De
w point t
emper
atur
e (°C)
-40
-20
0
20
20
-60
-80
0
10
30
-10
-20
-30
-50
-40
-70
-10
-30
10
30
40
50
60
70
-90
-100
Temperature of measured gas (°C)
Area not recommended
for continuous measurement
Accuracy not specified
Accuracy
±3 °C Td
Accuracy
±2 °C Td
Accuracy not specified
Figure 17 Accuracy Over Temperature Range
Table 39 Operating Environment
Property
Description/Value
Temperature
−40 … +70 °C (−40 … +158 °F)
Relative humidity
0 … 100 %RH (up to +20 °C / +68 °F)
Pressure
0 … 50 bar (725 psi
a
)
Measured gases
Non-corrosive gases
Sample flow rate
No effect on measurement accuracy
EMC compliance
EN61326-1, Industrial environment
Table 40 Inputs and Outputs
Property
Description/Value
Two analog outputs (scalable)
4 … 20 mA, 0 … 20 mA (3-wire) 0 … 5 V, 0 … 10 V
Digital output
RS-485 (2-wire)
Alarm-level indication by analog signal
User selectable
Purge information
5 V, 10 V, 20 mA, or LED
Accuracy of analog outputs
±0.01 V / ±0.01 mA
ON/OFF Output Available Instead of 2nd Analog Output
Purge inactive
0 V or 4 mA
Purge active
5 V, 10 V, or 20 mA
OR exceeded T
d
or ppm limit
User selectable
Chapter 8 – Technical Data
61
Содержание Drycap DMT152
Страница 1: ...M210865EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT152...
Страница 8: ...DMT152 User Guide M210865EN J 6...
Страница 68: ...DMT152 User Guide M210865EN J 66...
Страница 70: ...DMT152 User Guide M210865EN J 68...
Страница 71: ......
Страница 72: ...www vaisala com...