5.7 Troubleshooting and Maintenance Commands
5.7.1 Display Currently Active Errors
Use the
ERRS
command to display the currently active transmitter error codes. Possible errors
and their causes are listed in
Table 29 ERRS command
Syntax
Description
ERRS
<cr>
Displays the list of currently active error codes.
Example:
errs
Voltage too low error
5.7.2 Test Analog Outputs
The operation of the analog outputs can be tested with the
ATEST
command by forcing the
output to a given value which can then be measured with calibrated multimeter. The command
uses the currently selected analog output types for the channels (voltage or current) so you
do not have to specify the unit.
Before testing, you can verify the currently set analog output modes using the
AMODE
command. After testing the output, give the
ATEST
command again to exit the test mode.
Table 30 ATEST command
Syntax
Description
ATEST [xxx yyy]
<cr>
Enter the test output values for channel 1 and
channel 2.
xxx
Channel 1 output value (V or mA)
yyy
Channel 2 output value (V or mA)
The output shows the test values of the analog channels as well as diagnostic
information that may be useful to Vaisala Service Center if there is a problem
with the analog outputs.
DMT152 User Guide
M210865EN-J
48
Содержание Drycap DMT152
Страница 1: ...M210865EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT152...
Страница 8: ...DMT152 User Guide M210865EN J 6...
Страница 68: ...DMT152 User Guide M210865EN J 66...
Страница 70: ...DMT152 User Guide M210865EN J 68...
Страница 71: ......
Страница 72: ...www vaisala com...