![Vaisala Drycap DMT152 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362052.webp)
Syntax
Description
Example:
aover on
: ON
The following example illustrates how the analog output is affected.
Channel 1 outputs T
d
with voltage output 0 … 5 V (-80 °C … -30 °C). After giving the
AOVER ON
command, the range is 0 … 5.5 V (-80 °C … -25 °C).
Note that the -30 °C T
d
point is still at 5 V.
5.8 Other Commands
5.8.1 Opening the Transmitter in
POLL
Mode
Use the
OPEN
command to connect to a transmitter that is in
POLL
mode.
Table 33 OPEN command
Syntax
Description
OPEN [aa]
<cr>
[aa]
= address of the transmitter to connect to,
range: 0 … 99.
Example:
open 1
DMT152 1 line opened for operator commands
5.8.2 Closing the Connection to a Transmitter in
POLL
Mode
Use the
CLOSE
command to close a connection to a transmitter that is in
POLL
mode.
Table 34 CLOSE command
Syntax
Description
CLOSE
<cr>
Closes the transmitter connection.
Example:
close
line closed