User's Guide _______________________________________________________________________
8 ____________________________________________________________________ M210397EN-D
0507-094
Figure 2
Mounting Dewpoint Transmitter DMT142
The numbers in the following list refer to Figure 2 above:
1.
Select a proper location to mount the DMT142 and remove the
yellow sensor protective cap.
2.
Fit the sealing washer onto the probe.
3.
Mount the transmitter to the measurement point. The transmitter
has parallel thread G1/2" ISO 228/1.
4.
Connect the wires of the connection cable. When using the cable
provided with DMT142 refer to the following wiring section.
See the power supply requirements on page 11.
5.
Plug in the cable to the transmitter and turn on the power supply.
6.
Self diagnostics performed at start-up freeze the output during
the first minutes of operation. Typical warm-up time is about 5
minutes before normal operation.