User's Guide _______________________________________________________________________
26 ___________________________________________________________________ M210397EN-D
Serial Line Output Commands
Start Measurement Output
The transmitter outputs continuous measurement values as an ASCII
text string to the PC.
Syntax: R<cr>
Example:
>r
58.289 23.051 14.424
58.276 23.058 14.426
58.278 23.035 14.406
. . .
where
in the first column: relative humidity (%RH)
in the second column: ambient temperature (°C or °F)
in the third column: dewpoint (°C or °F)
Outputting the results continues in intervals issued with the command
INTV. Stop the output with the command S.
Stop Measurement Output
Syntax: S<cr>
Set Output Interval
Syntax: INTV n xxx<cr>
where
n
=
time interval = 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
Example:
>INTV 10 MIN
Output interval: 10 MIN
>
The shortest output interval (with n = 0 [zero]) is approximately one
second due to the internal measurement cycle.